Changes for page Testing Procedures

Last modified by robert on 2024/12/09 16:08

From version 9.1
edited by Jack Dent
on 2024/12/09 10:18
Change comment: There is no comment for this version
To version 5.2
edited by robert
on 2024/12/03 09:36
Change comment: There is no comment for this version

Summary

Details

Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.JackD
1 +XWiki.robert
Content
... ... @@ -26,7 +26,7 @@
26 26  
27 27  (% class="box infomessage" %)
28 28  (((
29 -NOTE: Unit powering ON is not instant, there may be a 10 to 15 sec delay.
29 +Unit powering ON is not instant, there may be a 10 to 15 sec delay.
30 30  )))
31 31  
32 32  
... ... @@ -39,7 +39,7 @@
39 39  
40 40  (% class="box infomessage" %)
41 41  (((
42 -NOTE: The following step, power feature is not present in LPR200, therefore this step cannot be verified for  LPR recorders.
42 +The following step, power feature is not present in LPR200, therefore this step cannot be verified for  LPR recorders.
43 43  )))
44 44  
45 45  == System Test ==
... ... @@ -67,7 +67,7 @@
67 67  
68 68  (% class="box infomessage" %)
69 69  (((
70 -NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.
70 +There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.
71 71  )))
72 72  
73 73  Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under.
... ... @@ -82,7 +82,7 @@
82 82  
83 83  (% class="box infomessage" %)
84 84  (((
85 -NOTE: If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.
85 +If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.
86 86  )))
87 87  
88 88  === Recording test ===
... ... @@ -103,7 +103,7 @@
103 103  
104 104  (% class="box infomessage" %)
105 105  (((
106 -NOTE: For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)
106 +For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)
107 107  )))
108 108  
109 109  === Troubleshooting ===
... ... @@ -454,18 +454,6 @@
454 454  
455 455  //(originally written by F. Bozinovic May 2024)//
456 456  
457 -
458 -=== Introduction ===
459 -
460 -Sensor calibration allows user to input an electrical test signal into a connected sensor to simulate ground motion. The resulting digitized sensor output can then be analysed to assess various attributes of the sensor, such as basic functionality,
461 -frequency response and/or sensitivity stability over time. For a high-quality broadband sensor, these parameters typically remain stable over time, so that if the sensor initially meets manufacturer’s specifications and has not suffered damage,
462 -then calibration is usually not required. However, calibration can be a useful quality control check if it is suspected that the sensor may be defective or damaged after multiple deployments.
463 -
464 -(% class="box infomessage" %)
465 -(((
466 -NOTE: Not all sensors have this capability and user must refer to the manufacturer’s datasheet for clarification.
467 -)))
468 -
469 469  === Process ===
470 470  
471 471  The Centaur data recorder can generate and output an analog signal using a 16-bit internal digital-to analog converter (DAC). The DAC output is applied to the sensor for calibration purposes via the matching sensor cable. Make sure to use manufacturer cables as the correct signal lines have been connected to the correct pins of the mating connector. The Centaur CTR, CTR2 and CTR3 series models may generate signals of up to ±5 V amplitude, while the Centaur CTR4 series models have an enhanced calibration output.
... ... @@ -624,25 +624,25 @@
624 624  
625 625  Inside the “Vrl (Theoretical)” cell enter the following formula using the corresponding cells.
626 626  
627 -V_{RL}=I_{oc}\times R_L
615 +[Equation]
628 628  
629 629  
630 630  Inside the “Rated Power” cell enter the following formula using the corresponding cells.
631 631  
632 -P_{oc}=V_{oc}\times I_{oc}
620 +[Equation]
633 633  
634 634  
635 635  Inside the “Load Power” cell enter the following formula using the corresponding cells.
636 636  
637 -P_{RL}=\frac{V_{RL}}{R_L}\times V_{oc}
625 +[Equation]
638 638  
639 639  
640 640  Inside the “Power Loss %” cell enter the following formula using the corresponding cells.
641 641  
642 -The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels.
630 +The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 
643 643  
644 644  
645 -Power\ Loss\ \%=\frac{P_{RL}}{P_{oc}}\times 100-100
633 +[Equation]
646 646  
647 647  
648 648  Perform all the calculations for each solar panel ID entered.