Changes for page Testing Procedures
Last modified by robert on 2024/12/09 16:08
Summary
-
Page properties (2 modified, 0 added, 0 removed)
Details
- Page properties
-
- Author
-
... ... @@ -1,1 +1,1 @@ 1 -XWiki. JackD1 +XWiki.robert - Content
-
... ... @@ -26,7 +26,7 @@ 26 26 27 27 (% class="box infomessage" %) 28 28 ((( 29 - NOTE:Unit powering ON is not instant, there may be a 10 to 15 sec delay.29 +Unit powering ON is not instant, there may be a 10 to 15 sec delay. 30 30 ))) 31 31 32 32 ... ... @@ -39,7 +39,7 @@ 39 39 40 40 (% class="box infomessage" %) 41 41 ((( 42 - NOTE: The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders.42 +The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders. 43 43 ))) 44 44 45 45 == System Test == ... ... @@ -67,7 +67,7 @@ 67 67 68 68 (% class="box infomessage" %) 69 69 ((( 70 - NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.70 +There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V. 71 71 ))) 72 72 73 73 Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under. ... ... @@ -82,7 +82,7 @@ 82 82 83 83 (% class="box infomessage" %) 84 84 ((( 85 - NOTE:If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.85 +If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test. 86 86 ))) 87 87 88 88 === Recording test === ... ... @@ -103,7 +103,7 @@ 103 103 104 104 (% class="box infomessage" %) 105 105 ((( 106 - NOTE:For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)106 +For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK) 107 107 ))) 108 108 109 109 === Troubleshooting === ... ... @@ -454,20 +454,8 @@ 454 454 455 455 //(originally written by F. Bozinovic May 2024)// 456 456 457 +=== Process === 457 457 458 -=== Introduction === 459 - 460 -Sensor calibration allows user to input an electrical test signal into a connected sensor to simulate ground motion. The resulting digitized sensor output can then be analysed to assess various attributes of the sensor, such as basic functionality, 461 -frequency response and/or sensitivity stability over time. For a high-quality broadband sensor, these parameters typically remain stable over time, so that if the sensor initially meets manufacturer’s specifications and has not suffered damage, 462 -then calibration is usually not required. However, calibration can be a useful quality control check if it is suspected that the sensor may be defective or damaged after multiple deployments. 463 - 464 -(% class="box infomessage" %) 465 -((( 466 -NOTE: Not all sensors have this capability and user must refer to the manufacturer’s datasheet for clarification. 467 -))) 468 - 469 -=== Process === 470 - 471 471 The Centaur data recorder can generate and output an analog signal using a 16-bit internal digital-to analog converter (DAC). The DAC output is applied to the sensor for calibration purposes via the matching sensor cable. Make sure to use manufacturer cables as the correct signal lines have been connected to the correct pins of the mating connector. The Centaur CTR, CTR2 and CTR3 series models may generate signals of up to ±5 V amplitude, while the Centaur CTR4 series models have an enhanced calibration output. 472 472 473 473 Calibration output signal actions are launched from the Waveform page in the Centaur Web interface. A synthetic waveform signal generator allows you to generate sinewave and pseudo-random binary (PRB) signals on demand. User can configure the sine frequency or PRB pulse width, signal duration and amplitude as well as specify lead-in and lead-out silence intervals before and after the calibration waveform. One can also select and play a calibration file containing any other desired digital time series waveform that by uploading it to the Centaur, such as a swept sinewave, step function, random noise, or chained PRB sequence. ... ... @@ -474,11 +474,11 @@ 474 474 475 475 The following sample calibration files are supplied with the Centaur. These files may be used to visually verify functionality and approximate sensitivity of the sensor by inspection of the output waveform: 476 476 477 -* **sine_5V_30s**generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds.478 -* **step_0V_to_5V_15s**generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds.479 -* **prb 1V 20ms 10min**generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude.480 -* **prb 1V 5s 150min**generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude.481 -* **prb 2V 5s 8hr**generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude.465 +* sine_5V_30s generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds. 466 +* step_0V_to_5V_15s generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds. 467 +* prb 1V 20ms 10min generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude. 468 +* prb 1V 5s 150min generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude. 469 +* prb 2V 5s 8hr generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude. 482 482 483 483 === Procedure === 484 484 ... ... @@ -624,25 +624,25 @@ 624 624 625 625 Inside the “Vrl (Theoretical)” cell enter the following formula using the corresponding cells. 626 626 627 - V_{RL}=I_{oc}\timesR_L615 +[Equation] 628 628 629 629 630 630 Inside the “Rated Power” cell enter the following formula using the corresponding cells. 631 631 632 - P_{oc}=V_{oc}\times I_{oc}620 +[Equation] 633 633 634 634 635 635 Inside the “Load Power” cell enter the following formula using the corresponding cells. 636 636 637 - P_{RL}=\frac{V_{RL}}{R_L}\times V_{oc}625 +[Equation] 638 638 639 639 640 640 Inside the “Power Loss %” cell enter the following formula using the corresponding cells. 641 641 642 -The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 630 +The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 643 643 644 644 645 - Power\ Loss\ \%=\frac{P_{RL}}{P_{oc}}\times100-100633 +[Equation] 646 646 647 647 648 648 Perform all the calculations for each solar panel ID entered.