Changes for page Testing Procedures
Last modified by robert on 2024/12/09 16:08
Summary
-
Page properties (2 modified, 0 added, 0 removed)
Details
- Page properties
-
- Author
-
... ... @@ -1,1 +1,1 @@ 1 -XWiki. JackD1 +XWiki.robert - Content
-
... ... @@ -26,7 +26,7 @@ 26 26 27 27 (% class="box infomessage" %) 28 28 ((( 29 - NOTE:Unit powering ON is not instant, there may be a 10 to 15 sec delay.29 +Unit powering ON is not instant, there may be a 10 to 15 sec delay. 30 30 ))) 31 31 32 32 ... ... @@ -39,7 +39,7 @@ 39 39 40 40 (% class="box infomessage" %) 41 41 ((( 42 - NOTE: The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders.42 +The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders. 43 43 ))) 44 44 45 45 == System Test == ... ... @@ -67,7 +67,7 @@ 67 67 68 68 (% class="box infomessage" %) 69 69 ((( 70 - NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.70 +There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V. 71 71 ))) 72 72 73 73 Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under. ... ... @@ -82,7 +82,7 @@ 82 82 83 83 (% class="box infomessage" %) 84 84 ((( 85 - NOTE:If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.85 +If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test. 86 86 ))) 87 87 88 88 === Recording test === ... ... @@ -103,7 +103,7 @@ 103 103 104 104 (% class="box infomessage" %) 105 105 ((( 106 - NOTE:For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)106 +For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK) 107 107 ))) 108 108 109 109 === Troubleshooting === ... ... @@ -454,7 +454,7 @@ 454 454 455 455 //(originally written by F. Bozinovic May 2024)// 456 456 457 -=== Process === 457 +=== Process === 458 458 459 459 The Centaur data recorder can generate and output an analog signal using a 16-bit internal digital-to analog converter (DAC). The DAC output is applied to the sensor for calibration purposes via the matching sensor cable. Make sure to use manufacturer cables as the correct signal lines have been connected to the correct pins of the mating connector. The Centaur CTR, CTR2 and CTR3 series models may generate signals of up to ±5 V amplitude, while the Centaur CTR4 series models have an enhanced calibration output. 460 460 ... ... @@ -462,11 +462,11 @@ 462 462 463 463 The following sample calibration files are supplied with the Centaur. These files may be used to visually verify functionality and approximate sensitivity of the sensor by inspection of the output waveform: 464 464 465 -* **sine_5V_30s**generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds.466 -* **step_0V_to_5V_15s**generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds.467 -* **prb 1V 20ms 10min**generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude.468 -* **prb 1V 5s 150min**generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude.469 -* **prb 2V 5s 8hr**generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude.465 +* sine_5V_30s generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds. 466 +* step_0V_to_5V_15s generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds. 467 +* prb 1V 20ms 10min generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude. 468 +* prb 1V 5s 150min generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude. 469 +* prb 2V 5s 8hr generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude. 470 470 471 471 === Procedure === 472 472 ... ... @@ -612,25 +612,25 @@ 612 612 613 613 Inside the “Vrl (Theoretical)” cell enter the following formula using the corresponding cells. 614 614 615 - V_{RL}=I_{oc}\timesR_L615 +[Equation] 616 616 617 617 618 618 Inside the “Rated Power” cell enter the following formula using the corresponding cells. 619 619 620 - P_{oc}=V_{oc}\times I_{oc}620 +[Equation] 621 621 622 622 623 623 Inside the “Load Power” cell enter the following formula using the corresponding cells. 624 624 625 - P_{RL}=\frac{V_{RL}}{R_L}\times V_{oc}625 +[Equation] 626 626 627 627 628 628 Inside the “Power Loss %” cell enter the following formula using the corresponding cells. 629 629 630 -The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 630 +The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 631 631 632 632 633 - Power\ Loss\ \%=\frac{P_{RL}}{P_{oc}}\times100-100633 +[Equation] 634 634 635 635 636 636 Perform all the calculations for each solar panel ID entered.