... ... @@ -26,7 +26,7 @@
26 26  
27 27  (% class="box infomessage" %)
28 28  (((
29 -Unit powering ON is not instant, there may be a 10 to 15 sec delay.
29 +NOTE: Unit powering ON is not instant, there may be a 10 to 15 sec delay.
30 30  )))
31 31  
32 32  
... ... @@ -39,7 +39,7 @@
39 39  
40 40  (% class="box infomessage" %)
41 41  (((
42 -The following step, power feature is not present in LPR200, therefore this step cannot be verified for  LPR recorders.
42 +NOTE: The following step, power feature is not present in LPR200, therefore this step cannot be verified for  LPR recorders.
43 43  )))
44 44  
45 45  == System Test ==
... ... @@ -67,7 +67,7 @@
67 67  
68 68  (% class="box infomessage" %)
69 69  (((
70 -There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.
70 +NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V.
71 71  )))
72 72  
73 73  Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under.
... ... @@ -82,7 +82,7 @@
82 82  
83 83  (% class="box infomessage" %)
84 84  (((
85 -If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.
85 +NOTE: If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test.
86 86  )))
87 87  
88 88  === Recording test ===
... ... @@ -103,7 +103,7 @@
103 103  
104 104  (% class="box infomessage" %)
105 105  (((
106 -For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)
106 +NOTE: For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK)
107 107  )))
108 108  
109 109  === Troubleshooting ===
... ... @@ -454,7 +454,7 @@
454 454  
455 455  //(originally written by F. Bozinovic May 2024)//
456 456  
457 -=== Process ===
457 +=== Process ===
458 458  
459 459  The Centaur data recorder can generate and output an analog signal using a 16-bit internal digital-to analog converter (DAC). The DAC output is applied to the sensor for calibration purposes via the matching sensor cable. Make sure to use manufacturer cables as the correct signal lines have been connected to the correct pins of the mating connector. The Centaur CTR, CTR2 and CTR3 series models may generate signals of up to ±5 V amplitude, while the Centaur CTR4 series models have an enhanced calibration output.
460 460  
... ... @@ -462,11 +462,11 @@
462 462  
463 463  The following sample calibration files are supplied with the Centaur. These files may be used to visually verify functionality and approximate sensitivity of the sensor by inspection of the output waveform:
464 464  
465 -* sine_5V_30s generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds.
466 -* step_0V_to_5V_15s generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds.
467 -* prb 1V 20ms 10min generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude.
468 -* prb 1V 5s 150min generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude.
469 -* prb 2V 5s 8hr generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude.
465 +* **sine_5V_30s** generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds.
466 +* **step_0V_to_5V_15s** generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds.
467 +* **prb 1V 20ms 10min** generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude.
468 +* **prb 1V 5s 150min** generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude.
469 +* **prb 2V 5s 8hr** generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude.
470 470  
471 471  === Procedure ===
472 472  
... ... @@ -612,25 +612,25 @@
612 612  
613 613  Inside the “Vrl (Theoretical)” cell enter the following formula using the corresponding cells.
614 614  
615 -[Equation]
615 +V_{RL}=I_{oc}\times R_L
616 616  
617 617  
618 618  Inside the “Rated Power” cell enter the following formula using the corresponding cells.
619 619  
620 -[Equation]
620 +P_{oc}=V_{oc}\times I_{oc}
621 621  
622 622  
623 623  Inside the “Load Power” cell enter the following formula using the corresponding cells.
624 624  
625 -[Equation]
625 +P_{RL}=\frac{V_{RL}}{R_L}\times V_{oc}
626 626  
627 627  
628 628  Inside the “Power Loss %” cell enter the following formula using the corresponding cells.
629 629  
630 -The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 
630 +The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels.
631 631  
632 632  
633 -[Equation]
633 +Power\ Loss\ \%=\frac{P_{RL}}{P_{oc}}\times 100-100
634 634  
635 635  
636 636  Perform all the calculations for each solar panel ID entered.