Changes for page Testing Procedures
Last modified by robert on 2024/12/09 16:08
Summary
-
Page properties (2 modified, 0 added, 0 removed)
-
Attachments (0 modified, 1 added, 0 removed)
Details
- Page properties
-
- Author
-
... ... @@ -1,1 +1,1 @@ 1 -XWiki. robert1 +XWiki.JackD - Content
-
... ... @@ -26,7 +26,7 @@ 26 26 27 27 (% class="box infomessage" %) 28 28 ((( 29 -Unit powering ON is not instant, there may be a 10 to 15 sec delay. 29 +NOTE: Unit powering ON is not instant, there may be a 10 to 15 sec delay. 30 30 ))) 31 31 32 32 ... ... @@ -39,7 +39,7 @@ 39 39 40 40 (% class="box infomessage" %) 41 41 ((( 42 -The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders. 42 +NOTE: The following step, power feature is not present in LPR200, therefore this step cannot be verified for LPR recorders. 43 43 ))) 44 44 45 45 == System Test == ... ... @@ -67,7 +67,7 @@ 67 67 68 68 (% class="box infomessage" %) 69 69 ((( 70 -There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V. 70 +NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V. 71 71 ))) 72 72 73 73 Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under. ... ... @@ -82,7 +82,7 @@ 82 82 83 83 (% class="box infomessage" %) 84 84 ((( 85 -If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test. 85 +NOTE: If the user requires to verify the sensor or the validity of the recorded data, a power spectral density analysis would need to be performed, see rest of document (link here) for instructions on how to perform this test. 86 86 ))) 87 87 88 88 === Recording test === ... ... @@ -103,7 +103,7 @@ 103 103 104 104 (% class="box infomessage" %) 105 105 ((( 106 -For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK) 106 +NOTE: For procedure on how to use and set-up the PSD script refer to the “Performing PSD function on recorded sensor data procedure” document. (LINK) 107 107 ))) 108 108 109 109 === Troubleshooting === ... ... @@ -295,7 +295,6 @@ 295 295 * Alternatively the SD card may be faulty and need to be replaced 296 296 ))) 297 297 298 - 299 299 (% class="table-condensed" style="width:1063px" %) 300 300 |(% colspan="2" style="width:1060px" %)((( 301 301 **General System Errors ** ... ... @@ -412,7 +412,6 @@ 412 412 * There is a damaged cell which will need to be replaced 413 413 ))) 414 414 415 - 416 416 (% class="table-condensed" style="width:1064px" %) 417 417 |(% colspan="2" style="width:1061px" %)((( 418 418 **Hardware Errors ** ... ... @@ -454,8 +454,52 @@ 454 454 455 455 == ... via Centaur (Nanometrics sensors) == 456 456 457 - GUIDETOTESTVIACENTAUR455 +//(originally written by F. Bozinovic May 2024)// 458 458 457 + 458 +=== Introduction === 459 + 460 +Sensor calibration allows user to input an electrical test signal into a connected sensor to simulate ground motion. The resulting digitized sensor output can then be analysed to assess various attributes of the sensor, such as basic functionality, 461 +frequency response and/or sensitivity stability over time. For a high-quality broadband sensor, these parameters typically remain stable over time, so that if the sensor initially meets manufacturer’s specifications and has not suffered damage, 462 +then calibration is usually not required. However, calibration can be a useful quality control check if it is suspected that the sensor may be defective or damaged after multiple deployments. 463 + 464 +(% class="box infomessage" %) 465 +((( 466 +NOTE: Not all sensors have this capability and user must refer to the manufacturer’s datasheet for clarification. 467 +))) 468 + 469 +=== Process === 470 + 471 +The Centaur data recorder can generate and output an analog signal using a 16-bit internal digital-to analog converter (DAC). The DAC output is applied to the sensor for calibration purposes via the matching sensor cable. Make sure to use manufacturer cables as the correct signal lines have been connected to the correct pins of the mating connector. The Centaur CTR, CTR2 and CTR3 series models may generate signals of up to ±5 V amplitude, while the Centaur CTR4 series models have an enhanced calibration output. 472 + 473 +Calibration output signal actions are launched from the **Waveform** page in the Centaur Web interface. A synthetic waveform signal generator allows you to generate sinewave and pseudo-random binary (PRB) signals on demand. User can configure the sine frequency or PRB pulse width, signal duration and amplitude as well as specify lead-in and lead-out silence intervals before and after the calibration waveform. One can also select and play a calibration file containing any other desired digital time series waveform that by uploading it to the Centaur, such as a swept sinewave, step function, random noise, or chained PRB sequence. 474 + 475 +The following sample calibration files are supplied with the Centaur. These files may be used to visually verify functionality and approximate sensitivity of the sensor by inspection of the output waveform: 476 + 477 +* **sine_5V_30s** generates a 1 Hz sine wave with 5 V amplitude lasting 30 seconds. 478 +* **step_0V_to_5V_15s** generates a 0 V signal for 15 seconds followed by a positive 5 V step function lasting 15 seconds. 479 +* **prb 1V 20ms 10min** generates a 10 minute PRB sequence using 20 ms pulses and 1 V amplitude. 480 +* **prb 1V 5s 150min** generates a 2.5 hour PRB sequence using 5 second pulses and 1 V amplitude. 481 +* **prb 2V 5s 8hr** generates an 8 hour PRB sequence using 5 s pulses and 2 V amplitude. 482 + 483 +=== Procedure === 484 + 485 +1. Log-in to the Centaur Web Interface and use the Admin credentials 486 +1. Navigate to the **Health** page and verify that the sensor is properly levelled and recognised by its serial number. 487 +1. To configure the calibration parameters, navigate to the **Waveform** page. 488 +1. From the Calibration panel at the top page, select **Type** from the drop-down list and choose Sine. 489 +1. For the CTR4 series models, additional option to select between Voltage or Current is available. 490 +1. Click on the **Configure** button to access the calibration dialog box for the selected **Playback**. 491 +1. Configure the signal characteristics by selecting 5V, 30 sec with gain of 1. 492 +1. Configure the padding before and after the calibration signal, enter 5 seconds. 493 +1. The **Duration (s)** time can be made shorter or longer as required by user. NOTE, for shorter frequencies a longer duration will be required for the signal to complete its full cycle and to capture the entire waveform on the screen. 494 +1. Click OK button to close the dialog box and save the settings. 495 +1. Click the start calibration button [[image:1733178329484-829.png]] to begin the process. Approximately 5 seconds of time padding ( as set in Step 8) will past before the sensor responds to the injected signal and display the sine wave feedback response. 496 +1. The calibration will end after 30 seconds (as set in Step 7) or can be terminated manually by pressing the stop button. The calibration will then stop after 5 seconds and any configured lead out silence will be skipped. 497 +1. Click on the pause button on the bottom of the page to stop the live stream and use the arrows to centre the sine signal response. Note that the screen waveform will turn Blue and the stream will freeze. Avoid capturing live streaming signal! 498 +1. Once the response signal is cantered, perform a screen-capture “Print Screen” button or use “Snippet” tool and save the captured image locally. In the nametag, include a Serial number, date and sensor type. 499 +1. Archive and back-up the file. 500 + 459 459 == ... via data comparision (vs S1.AUANU) == 460 460 461 461 CODE & Guide to use code ... ... @@ -465,7 +465,7 @@ 465 465 466 466 //(originally written by F. Bozinovic November 2024)// 467 467 468 -Testing solar panels is vital for any remote seismic station, since role ofthe solar panel ensures that the batteries are kept charged throughout the day. Therefore, reliably testing them ensures only the working panels are installed on remote sites, ensuring success of the site operation and serviceability.510 +Testing solar panels is vital for any remote seismic station, since the solar panel ensures that the batteries are kept charged throughout the day. Therefore, reliably testing them ensures only the working panels are installed on remote sites, ensuring success of the site operation and serviceability. 469 469 470 470 This procedure describes the method for testing solar panels and determining how to identify defective panels. The testing of solar panels should be performed outdoors, under bring sun to obtain accurate results. 471 471 ... ... @@ -475,7 +475,7 @@ 475 475 ))) 476 476 477 477 478 -Following materials are required 520 +=== Following materials are required === 479 479 480 480 * Solar panel for testing 481 481 * Digital multi-meter (DMM) ... ... @@ -485,7 +485,7 @@ 485 485 * Spreadsheet with formulae 486 486 * Marker/ pen 487 487 488 -Test Method 530 +=== Test Method === 489 489 490 490 1. Clearly label each solar panel to keep track of measurements. 491 491 1. Record the manufacturers power rating of the solar panel. **Perform all measurement outdoors under bright sunny conditions! ** ... ... @@ -500,7 +500,7 @@ 500 500 501 501 502 502 503 -Developing a spreadsheet 545 +=== Developing a spreadsheet === 504 504 505 505 Create a spreadsheet with following cells 506 506 ... ... @@ -582,25 +582,25 @@ 582 582 583 583 Inside the “Vrl (Theoretical)” cell enter the following formula using the corresponding cells. 584 584 585 - [Equation]627 +V_{RL}=I_{oc}\times R_L 586 586 587 587 588 588 Inside the “Rated Power” cell enter the following formula using the corresponding cells. 589 589 590 - [Equation]632 +P_{oc}=V_{oc}\times I_{oc} 591 591 592 592 593 593 Inside the “Load Power” cell enter the following formula using the corresponding cells. 594 594 595 - [Equation]637 +P_{RL}=\frac{V_{RL}}{R_L}\times V_{oc} 596 596 597 597 598 598 Inside the “Power Loss %” cell enter the following formula using the corresponding cells. 599 599 600 -The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 642 +The calculated values that are negative represent power loss, and positive values are power gain. Performing “conditional formatting” on these cells with colour gradient (defined by colour break limits) would yield visually easy to recognise defective panels. 601 601 602 602 603 - [Equation]645 +Power\ Loss\ \%=\frac{P_{RL}}{P_{oc}}\times 100-100 604 604 605 605 606 606 Perform all the calculations for each solar panel ID entered.
- 1733178329484-829.png
-
- Author
-
... ... @@ -1,0 +1,1 @@ 1 +XWiki.robert - Size
-
... ... @@ -1,0 +1,1 @@ 1 +1.5 KB - Content