Changes for page Testing Procedures
Last modified by robert on 2024/12/09 16:08
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -65,9 +65,9 @@ 65 65 66 66 Set-up an external power source, by connecting a solar regulator to a power supply and set voltage to 18V DC. Verify the regulator is supplying the correct voltage of 7.7V and then plug into the “External Power” port of the recorder. 67 67 68 -(% class="box warningmessage" %)68 +(% class="box infomessage" %) 69 69 ((( 70 - **IMPORTANT!**TheLPR-200 now hastwo types of solar regulators.The new type (exclusiveafter Jan 1 2025) are modified to output 13.8V (for chargingleadacid batteriesdirectly).Do notuse these withtheold lithiumbattery packsasthismay destroy theunitsand possiblyexplode!Ifyouare confusedemail us.70 +NOTE: There are two types of solar regulators available. Make sure to not mix them as they supply different Voltages, 7.7V and 13.8V. 71 71 ))) 72 72 73 73 Navigate to “System Information” in the Menu and note the battery icon will have a lightning symbol indicating it’s charging, also observe the state of charge % is increasing under.