Wiki source code of SmartSolo Node Seismometers

Version 117.2 by robert on 2026/08/12 12:56

Show last authors
1 (% class="row" %)
2 (((
3 (% class="col-xs-12 col-sm-8" %)
4 (((
5 = **Node Types** =
6
7 ANSIR supply two types of three-channel nodes, and one type of one-channel node:
8
9 * **SmartSolo IGU-16HR 3C (5 Hz, 'very' short period)**
10 * **SmartSolo BD3C-5 (5 second, short period)**
11 * **SmartSolo IGU-16 1C (5 Hz, 'very' short period, single channel. Not 'HR')**
12
13 Visit the [[SmartSolo page>>https://smartsolo.com/igu.html]] for more detail.
14
15 The three-channel nodes have a theoretical battery capacity of ~~30 days, whereas the single-channel type has a capacity of ~~50 days. The programming, operation and downloading procedures for all types of SmartSolo nodes are also similar.
16
17 (% class="box infomessage" %)
18 (((
19 **Freight update, 2026: **Freight options for lithium-ion batteries are changing in 2025/2026 to comply with updated transport safety regulations. This will impact supply of IGU 16 (<100Wh) and BD3C (168Wh) nodes. Advice will be sought from freighters on a case-by-case basis while they implement new guidelines.
20 )))
21
22 ----
23
24 = **Programming Defaults** =
25
26 The nodes must be programmed in the SoloLite software prior to use. The screenshots below show our recommended parameters for the 5 Hz (16HR-3C) and 5 second (BDC3-5) nodes.
27
28 [[IGU16HR-3C programming screen set at 250 Hz. Ensure circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]]
29
30 [[BD3C-5 programming screen set at 250 Hz. Ensure circled areas are set!>>image:5S_node_programming.labels.png||alt="BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!"]]
31
32
33 FIFO (first in, first out) data mode is safest as this will overwrite old data in case you forgot to clear the storage. At <= 250 hz you can fit 4++ months of data on these, shouldn't be an issue.
34
35 Note that the samplerate is instead given in sample spacing, in milliseconds. 4 ms = 250 Hz, 1 ms = 1000 Hz, 10 ms = 100 Hz, ad nauseam.
36
37 Be sure to set the channel types to Seismic and the gain appropriately. For active source (i.e. explosions) you can leave the gain at 0, but for passive experiments some gain is purported to be helpful (although we have found this to be somewhat negligible). We can confirm that 6db for the broadband nodes and 24db for the short period works well.
38
39 If you want recording to begin immediately, ensure that begin date is in the past by at least a few days. If you prefer to have a timed turn-on, then set the times as needed. **Be warned that recording will not begin until a GPS lock is achieved, so if you set it to a future turn-on but bury it too deeply, it may not record!**
40
41 In newer versions of the software there is a "power consumption" setting.. we are no sure what this does exactly, but the manual says it gives life a "boost" at the expense of about 3 db of resolution. Until this can be quantified a bit better it is not recommended.
42
43 Storage type can be DLD (proprietary) or Miniseed. We assume the software works best with their proprietary format so prefer not to risk any issues. You can export to miniseed later.
44
45 GPS is best set to cycle mode (e.g. once per hour) instead of constant "always on". The clock drift on these are almost nil even if there is no sync at all, so it's best to conserve power.
46
47 Bluetooth (BD3C-5 only) should be turned OFF to conserve power.
48
49 We recommend that the 16HR-3C be set to a gain of 24db for passive experiments and no higher than 250 Hz sampling rate unless there is an explicit reason to do so. The BD3C-5 should be set to a gain of 6db (which is the maximum allowed) for passive experiments (or 0 db if active).
50
51 {{info}}
52 **Note that any applied instrument gain must be removed when exporting (e.g. to miniseed) after your deploy, **otherwise amplitudes will be a factor of either 15.84893192 (24db) or 2 (6db) too high!
53 {{/info}}
54
55 = **Fieldwork Preparation** =
56
57 (% class="box warningmessage" %)
58 (((
59 **INVEST IN FAST EXTERNAL HARD DRIVES – DO NOT LET THIS BE THE LIMITATION OF DATA HARVESTING**
60
61 **Assume ~~1 Tb of storage for both raw and exported data per 50 nodes @ 250 Hz & 30 days. One node recording at 250 Hz for 30 days tends to create about 3 Gb of miniseed data.**
62
63 **We have had good experience with the 4Tb Samsung T7 Shield drives.**
64 )))
65
66 == Magnets ==
67
68 If you are short on magnets, you may find it easier and a lot cheaper to buy magnets in Australia. AMF Magnetics is a good retailer, and [[this item>>https://magnet.com.au/collections/shop?q=23012B]] seems to work well. It is also advantageous to use smaller magnets and store/carry them individually in your back pocket (as well as stick them to various places in your field vehicle, etc).
69
70 == Animal-Proofing ==
71
72 We have experienced interference from animals (foxes, dogs, goats) digging up and carrying nodes off for tens or hundreds of metres. It is helpful to minimise human and food smells (particularly on the rope handles) when working in areas where this is a risk. Or, wipe down affected nodes with 50-80% methylated spirits if extensive handling can not be avoided.
73
74 If you come to collect your node and it is missing~-~- LOOK FOR IT! It may not have gotten far. We have found dozens of nodes by spending 15 minutes looking for them.
75
76 = External Power =
77
78 Both the 5Hz IGU-16HR and 5s BD3C-5 can be optionally powered via external battery via either a replacement bottom half (the 5Hz nodes) or a battery cable accessory (BD3C-5) using standard lead acid batteries from 9-36v. We have done preliminary testing at 250 Hz with a 12v battery:
79
80 - BD3C-5: ~~2 days of recording per 1 Ah
81
82 - IGU16-HR 3C:  ~~3.5 days of recording per 1 Ah
83
84 - IGU16-HR 1C: ~~7 days of recording per 1 Ah
85
86 Of course, these can also be fit with a solar panel & charge controller which would then theoretically keep them going indefinitely (limited only by the disk storage, which at 250 Hz could be on the order of 12 months).
87
88 = **Installation** =
89
90 (% class="box infomessage" %)
91 (((
92 **Field logs are a critical component of fieldwork and this is especially the case for large N nodal deploys. Take notes!**
93 )))
94
95 == 1. Logbook documentation ==
96
97 (((
98 **Essential Details** for field logs:
99
100 * Station name
101 * Latitude, longitude, elevation
102 * Names of team members present
103 * Date and both local & UTC time of installation/removal
104 * Serial number (SN) of the TOP HALF of the sensor (if a BD3C-5, there is only one serial number)
105 * Detailed notes on the site conditions and setup, anything else that will be helpful to find it again ("by the fence", "south of rock", etc)
106
107 [[HERE>>http://auspass.edu.au/field/NODES_blank_fieldlog.pdf]] is an example logsheet that works well for nodes, feel free to print and use!
108
109 == 2. Burial ==
110
111 ALL nodes should be buried flush with the ground, no exceptions. If the node is sticking up the data data quality will be significantly reduced (especially the horizontal channels, see [[below>>https://auspass.edu.au/xwiki/bin/view/Instrumentation/SmartSolo%20Nodes/#HIGU-16Horizontalnoise26howtoavoid]]). It also catches wind, makes them more visible, etc. But mostly it makes the data bad. Don't be lazy: do it right.
112
113 [[An example of what not to do.>>image:1786503344572-296.jpg]]
114
115 == 3. Node Placement ==
116 )))
117
118 **Site Analysis**:
119
120 * **Take compass measurements away from the sensor as it will affect your measurement. Use a stick or shovel to help align.**
121 * Take photographs from various angles to document the site setup thoroughly. Have a colleague stand next to it pointing at it.
122 * Include a detailed site description in your notes.
123
124 == 4. GPS Considerations ==
125
126 (% class="wikigeneratedid" %)
127 The GPS antenna is at the top and center of the unit, and will (usually) only receive signal with a clear sky view directly above. The signal is able to penetrate plastic and terracotta planters and a thin (2 cm?) layer of soil, but may struggle if the soil layer is too thick. **These nodes will not start recording without attaining a GPS lock** and repeated attempts will excessively drain the battery.
128
129 == 5. Visibility and Location Marking ==
130
131 **Flag Placement**: Position a flag, preferably in a bright color (avoid green or yellow), near the instrument to aid in its future location.
132
133 **GPS Marking**:
134
135 * Use a GPS device to mark the instrument's exact location. Most modern cell phones can get to about a 3m error with their internal GPS also; you can probably also get away with investing a few dollars in a good app that shows error and lets you log markers.
136 * Also write the GPS down on paper (ie your [[LOG SHEET>>http://auspass.edu.au/field/NODES_blank_fieldlog.pdf]]).
137
138 == (% style="color:inherit; font-family:inherit; font-size:max(18px, min(20px, 14.4444px + 0.462963vw))" %)6. Charge Time, Pre-Deployment & Post-Deployment(%%) ==
139
140 * **Charging Duration**: Both types of nodes take approximately 6-8 hours to fully charge from a flat state.
141 * **Pre-Deployment Charging**:
142 ** Although the nodes hold their charge well, it's beneficial to give them a "top up" charge before deployment.
143 * **Operational Duration**:
144 ** When recording at 250 Hz, with GPS on and Bluetooth disabled, the instruments are expected to last about 30 days per charge cycle. If they are set to run only overnight, this can be extended to 60 days.
145
146 * **Post-Retrieval Charging**:
147 ** After retrieval, charge the instruments to about 50-60% (indicated by ORANGE LED) unless they are to be immediately re-deployed or transported.
148 * **State of Charge (SoC) for Storage**:
149 ** Maintain a battery charge level of around 50-60% (i.e., ORANGE) for storage.
150 ** This charge level is recommended to prevent battery damage, and should be checked every six months.
151 ** Nodes should //__not be stored at full-charge (GREEN), or 0-charge (RED).__//
152 ** Storage at 0-charge damages lithium batteries**.**
153 * **SoC for Transport:**
154 ** Charge levels for transport will be advised by the freighter. The required SoC will depend on volume and transport method (air, land, sea).
155
156 (((
157 == 7. Data Sharing and Metadata Creation ==
158 )))
159
160 **GPS Data**:
161
162 * Ensure you have __carefully documented__ precise lat/lon locations for each station.
163
164 **Photo Sharing**:
165
166 * It is strongly encouraged to take pictures of each site and upload these to a shared platform (OneDrive, Dropbox, etc.).
167
168 **Metadata File**:
169
170 * Create and organize metadata according to the [[ANU metadata standard txt file>>attach:example_metadata.txt]]. This is going to be particularly important if you are reusing nodes at different sites... not documenting the serial numbers (of the **top half** of the node) and the times they were deployed can lead to station mix-ups.
171
172 == 8. Additional Best Practices ==
173
174 * **Training and Familiarisation**: Make sure all team members are adequately trained in using the GPS devices, compass use, and other equipment to ensure consistent and accurate data collection.
175
176 ----
177
178 = **Seismic Station Demobilization and Documentation** =
179
180 1. (((
181 **Preparation for Demobilization**:
182
183 * Before starting the demobilization process, ensure you have a compass, tape, marker, pen, masking tape, clipboard, logbook, and compass ready in your tote bag.
184 )))
185 1. (((
186 **Locating the instrument**:
187
188 * Use the downloaded GPS file to accurately locate the node for demobilization.
189 * Import this KMZ file onto your phone for easy reference and location tracking.
190 * Utilize Google Maps or Google Earth to create a KMZ file of the station’s location.
191 )))
192 1. (((
193 **Labeling Instruments for Demobilization**:
194
195 * Write the station name and the instrument’s serial number on a masking tape label to apply to the top of the node.
196 * Add markers 'D' (for download), 'C' (for charge), and ‘R’ (for removal) next to checkboxes on the label.
197 * Affix this label to the top of the instrument to avoid confusion during the charging and downloading data.
198 )))
199 1. (((
200 **Photographing the Setup Node**:
201
202 * Take a photo of the entire setup node with the __//label//__ and __//compass visible//__.
203 * This photo serves as a final record of the instrument’s condition and orientation at the time of removal.
204 )))
205 1. (((
206 **Logging Demobilization Details**:
207
208 * Use the field logbook to note the time of demobilization, serial numbers, and station name.
209 * Record any observations or issues related to the instrument’s orientation, level, or any other relevant factors.
210 )))
211 1. (((
212 **Final Checks and Equipment Removal**:
213
214 * Before physically removing the instrument, double-check that all necessary data has been downloaded and all photos and notes have been taken.
215 * Carefully dismantle and pack the equipment, ensuring that all components are accounted for and securely stored for transport.
216
217
218 )))
219
220 [[image:1706153556166-231.jpeg||data-xwiki-image-style-alignment="center" height="345" width="460"]]
221
222 ----
223
224 = **Charging Procedure for Seismic Nodes** =
225
226 (((
227 == 1. Preparation for Charging: ==
228
229 * Before charging, ensure each node is clean. This involves removing any dirt or debris to maintain the integrity of the equipment and ensure effective charging.
230 )))
231
232 (((
233 == 2. Disassembling the Node: ==
234
235 * For the IGU-16HR, remove the battery section (bottom half) from the sensor by unscrewing the spike section counter-clockwise.
236 )))
237
238 (((
239 == 3. Setting Nodes in the Charging Box: ==
240
241 * Connect to a safe indoor power supply, and turn on (red rocker switch).
242 * Charging will begin automatically when nodes are inserted in the charging rack.
243 * Place IGU-16HR battery sections upside-down in the rack, oriented with the terminal connectors.
244 )))
245
246 (((
247 == 4. Monitoring the Charging Process: ==
248
249 * Lights adjacent to the batteries will illuminate, indicating that charging is underway.
250 * Observe the transition of the lights from steady RED to ORANGE, then GREEN, and finally to FLASHING GREEN. A flashing green light indicates the batteries are fully charged.
251 )))
252
253 (((
254 == 5. Updating Charge Status: ==
255
256 * During the charging period, take this opportunity to update the status of each unit. Check the //"C"// box on your temporary labels to indicate that the unit has been successfully charged.
257 * This step is crucial for tracking the charging status of multiple units, especially when handling a large number of nodes.
258 )))
259
260 [[IGU 16-HRcharger (left) and harvester (right)>>image:1705195933422-337.png||data-xwiki-image-style-alignment="center" height="299" width="530"]]
261
262
263
264
265 {{{
266 }}}
267
268 ----
269
270 = **Downloading and Converting Seismic Data to MiniSeed Format** =
271
272 (% class="wikigeneratedid" %)
273 SmartSolo provides the following powerpoint for SmartSolo node programming and operation. Note that ANSIR only uses a portion of their process for our own uses: [[https:~~/~~/nappe.wustl.edu/smartsolo/files/smartsolo_online_training.pdf>>url:https://nappe.wustl.edu/smartsolo/files/smartsolo_online_training.pdf]]
274
275 == Connection tips: ==
276
277 Connecting SmartSolo nodes to their harvesters and having SoloLite recognise them can be a bit tricky, here are some tricks to help register and program them easier:
278
279 * Place the node on the harvester gently, then firmly press it down onto the pins.
280 * Place all nodes onto the harvester before trying any troubleshooting, as they may not show up while the SoloLite software is running. Once all are connected, try restarting the software for them to be recognised.
281 * If a node is refusing to connect, try it with another slot. It is easiest if you place all 16 nodes on the harvester, and swap any nodes that refuse to connect with each other.
282 * Nodes will likely not show up in the order that they should, though this is not an issue. E.g, a node in slot 6 on the harvester may show up in port 13 in the SoloLite software. Annoying, but it doesn't matter so long as you keep track of what's been harvested!
283
284
285
286 == Node Registration and Software Setup ==
287
288 1. (((
289 **Registering Nodes in the System**:
290
291 * To begin, register the nodes in the system so the software can recognize them.
292 * Navigate to the installation folder of “SmartSoloApps SoloLite”.
293 * Right-click on deviceconfig.exe and choose “run as an administrator”. Save the file to the “deviceconfig” directory (refer to the snapshot below).
294 * To avoid double registration, replace the file each time you register a new node.
295 )))
296 1. (((
297 **Creating a New Project in SoloLite**:
298
299 * Open the “SoloLite” software.
300 * Go to “File” and create a new project. Don't worry about finding the exact 16 nodes used in script writing.
301 * Ignore the settings for seismic recordings in the subsequent window. Resetting instruments (e.g., sampling rate, gain) requires reprogramming via script.
302 )))
303
304 == File structure ==
305
306 There are essentially three main folders where relevant PROSPECT and PROJECT DATA is stored. Individual projects will be found as subfolders in these.
307
308 === SOLOLITE ===
309
310 This folder stores SoloLite config files and parameters. Nothing too important stored here, you can always start over and re-create this.
311
312 === DCCDATA ===
313
314 This folder stores the RAW data you have harvested from the nodes. The data will still be on the nodes (unless you erased it) in case of emergency, but regardless, this is the folder you want to back up and save somewhere.
315
316 If you had a weird time harvesting a node, you can always manually copy it as if it were a USB stick and place it into this folder manually. The structure is: //C:/DCCDATA/prospect_name/project_name/SERIALNUMBER/label(usually a timestamp but can be anything)//
317
318 Then in the SoloLite software, go to tools > Reanalyze Seismic Data
319
320 === SOLODATA ===
321
322 This folder stores **exported** (e.g. miniseed) data. It is structured similarly. If your DCCDATA is intact, this can always be re-created if need be.
323
324
325 == Data Downloading Process ==
326
327 1. (((
328 **Initiating Data Download**:
329
330 * Once a new project is created, the Data Transfer View panel will display connected nodes with details like series number and data size.
331 * If “Prospect not matched” appears, it simply means the new project doesn’t match the original programming project. This is not a concern.
332 * Select all nodes and right-click to “force download”. This starts the download process.[[image:Smartsolo harvesting #4 copy.png]]
333 * Completed downloads will appear as new folders in the Downloaded Data panel.
334 )))
335 1. (((
336 **Exporting Data in Readable Format**:
337
338 * Go to the “Tool” menu and select “export seismic data”.
339 * Tailor other parameters to project preference and ensure "Sample Interval" matches the setting used during node reset (note: the standard used by ANU is 4ms, or 250hz)
340 * (% class="box warningmessage" %)
341 (((
342 **Ensure export data is set to "COUNTS" (int32), not "mV" (float). This is critical!**
343 )))
344 * Set "Remove Gain" to the same decibel gain as during programming. By default ANU sets this to 24db for short period nodes (a scaling factor of 15.848932), and 6db (a factor of 2.0) for broadband nodes.
345 * Set "Remove DC" to "Yes" to centre the data around the zero value
346 * Set the correct Start Time (UTC) and End Time (UTC) of the project to prevent the unnecessary export of older data
347 * [[image:Smartsolo harvesting #9 copy.png]]
348 * Click “prepare” followed by “run” to start reformatting. Monitor this process in the small panel at the bottom left.
349 * The data will be exported to the SOLODATA folder. For a windows system, the following file explorer page is where you must navigate to to locate your project folder[[image:Smartsolo harvesting #8 copy.png]]
350 )))
351
352 == Smart Solo IGU-16HR Polarity Notice ==
353
354 See [[5Hz Node Polarity Issues>>https://auspass.edu.au/xwiki/bin/view/Data/AusPass%20Data/#HSmartSoloNodePolarityIssues]] for discussion. If data is headed to AusPass, we prefer to invert the IGU-16HR channel data manually rather than in the SoloLite software or inverting the response metadata.
355
356 **The BD3C-5 data does not require any sort of polarity inversion.**
357
358 == 18 Leap Second bug ==
359
360 Not so much a //bug// as much as "a thing that can happen if your SoloLite installation is corrupted". If you notice your data has large constant time offsets, you should suspect that the number of leap seconds has not been accounted properly. There is a file "smartsoloconfig.xml" that needs to be present in "C:\SmartSoloApps SoloLite" (e.g. the main program directory) that dictates the leap second offset for the last two data ranges. Since 2017-01-01, this is 18 seconds. At some point in the next few years it will be 19 seconds.
361
362 If this file is missing, just create a new one structured like so, name it "smartsoloconfig.xml" and put it in your main program directory. Then, Reanalyze your data (tools > Reanalyze seismic data) and your data should have the correct time. You can also do this manually, if you want. The offset is 18 seconds precisely.
363
364 {{code language="none"}}
365 <?xml version="1.0" encoding="UTF-8"?>
366 <config>
367 <leapsecond>
368 <interval>
369 <start_time>2017-01-01#00:00:00</start_time>
370 <end_time>2999-12-31#23:59:59</end_time>
371 <second>18</second>
372 </interval>
373 <interval>
374 <start_time>1970-01-01#00:00:00</start_time>
375 <end_time>2017-01-01#00:00:00</end_time>
376 <second>17</second>
377 </interval>
378 </leapsecond>
379 <GPS_distance_threshold_degree>
380 4e-5
381 </GPS_distance_threshold_degree>
382 </config>
383 {{/code}}
384
385
386 == Handling Nodes During Download ==
387
388 1. (((
389 **Monitoring Download Indicators**:
390
391 * During download, green lights on nodes will blink, and associated red lights on the rack will flash.
392 * Disconnect nodes properly before unplugging anything.
393 * Be cautious: if the laptop enters sleep mode, the download will pause.
394 )))
395 1. (((
396 //**Investment in Storage Hardware**~://
397
398 * (% class="box warningmessage" %)
399 (((
400 * **Use fast external hard drives to avoid limitations in data harvesting.**
401
402 * **Recommended specifications: USB-C, USB 3.0, and 4+ Tb of space.**
403 * **The USB type for the harvester is TYPE-A, the typical normal rectangular shape.**
404 )))
405 )))
406 1. (((
407 **Metadata and Time Settings**:
408
409 * Ensure all metadata is saved with the file.
410 * System auto-determines the earliest data time as the start time. You can set it a day earlier at 00:00:00 for 24-hour data segments starting from midnight.
411 )))
412 1. (((
413 **Finalizing the Download**:
414
415 * After downloading, mark the //"D"// box on your temporary labels to indicate completion.
416 )))
417
418
419
420
421 [[Caption>>image:1705195543887-977.png||data-xwiki-image-style-alignment="center" height="534" width="632"]]
422
423
424
425
426 [[Caption>>image:1705195543890-537.png||data-xwiki-image-style-alignment="center" height="397" width="665"]]
427
428
429
430 [[Caption>>image:1705195543891-334.png||data-xwiki-image-style-alignment="center" height="379" width="650"]]
431
432
433
434 [[image:1705195543898-365.png||data-xwiki-image-style-alignment="center" height="467" width="674"]]
435
436 ----
437
438 = Instrument Response =
439
440 We are aware that there are various different published responses for these instruments and trust very few of them. One has to be careful with how polarity is handled between groups as well, and if one is working in integer counts (the ANSIR default) or mV (unclear why anyone would use this as it makes file sizes enormous). The response information published below is in **counts** and seems to fit well in huddle tests. Note that the response is the same for all channels and all units (e.g. there are no bespoke calibrations!), all appear to be sample rate insensitive, and the IGU data has been inverted (multiplied by -1) as described here: [[5Hz Node Polarity Issues>>https://auspass.edu.au/xwiki/bin/view/Data/AusPass%20Data/#HSmartSoloNodePolarityIssues]]
441
442 == IGU 16HR-3C ==
443
444 '16HR3C': {'poles':[(-22.211059+22.217768j), (-22.211059-22.217768j)],
445 'zeros':[0j, 0j],
446 'gain':1,
447 'sensitivity': 257019225.55108312}
448
449 [[X axis is samples (.01 s), Y axis is velocity (m/s), 0.5-5 Hz filter>>image:16HR_Z_huddle.png||alt="IGU16_Z_huddle.png"]]
450
451 [[X axis is samples (.01 s), Y axis is velocity (m/s), 0.5-5 Hz bandpass filter>>image:16HR_N_huddle.png||alt="IGU16_N_huddle.png"]]
452
453 == IGU 16-1C ==
454
455 The 1C nodes seem to have the same response as the 3-channel IGU-16HR-3C (above), however the response posted at IRIS-NRL seems to imply that there is no poles and zeros information (e.g. a flat/linear response). This is 100% not so.
456
457 [[IGU-16 1C, X axis is samples (.01 s), Y axis is velocity (m/s), 0.5-5 Hz bandpass filter. Seems to be same response as IGU-16HR-3C.>>image:16HR1C_Z_huddle.png||alt="IGU16_1C_Z_huddle.png"]]
458
459 == BD3C-5 ==
460
461 'BD3C': {'poles':[(-1720.4+0j), (-1.2+0.9j), (-1.2-0.9j)],
462 'zeros':[(14164+0j), (-7162+0j), 0j, 0j],
463 'gain':1.69726e-05,
464 'sensitivity': 702651512.6046528}
465
466 Above 0.5 Hz, the BD3C-5 response fits well:
467
468 [[X axis is samples (.01 s), Y axis is velocity (m/s), 0.5-5 Hz filter>>image:BD3C_Z_huddle.0.5.png]]
469
470 [[X axis is samples (.01 s), Y axis is velocity (m/s), 0.5-5 Hz bandpass filter>>image:BD3C_N_huddle.0.5.png]]
471
472 (% class="wikigeneratedid" %)
473 Below the corner frequency (0.2 Hz) the phase response still fares well, but amplitude response may need to be dialed in a bit (it seems a bit high). We are working to try to calibrate this a bit better. In the next two figures the filter is **0.1** to 5 Hz:
474
475
476 [[BD3C **0.1 **to 5 Hz bandpass filter>>image:BD3C_Z_huddle.0.1.png]]
477
478 [[BD3C **0.1** to 5 Hz bandpass filter>>image:BD3C_N_huddle.0.1.png]]
479
480 == IGU-16 Horizontal noise & how to avoid ==
481
482 The** 5 Hz nodes** are susceptible to horizontal noise due to the placement of geophones in the units, **but this can be mitigated by completely burying the units flush with the ground.** In the below example, the node was set on the floor of our basement set on its plastic carrying case support. As such the amount of horizontal noise noticeably increases above ~~ 10Hz.
483
484 [[IGU-16HR-3C Power spectrum huddle test vs a CMG-6TD (S1) and TC120/Centaur combo. The N and E channels have excess noise above 10Hz due to "sticking up" out of the ground.>>image:IGU16_spectrum.png]]
485
486 (% class="wikigeneratedid" %)
487 The BD3C-5 nodes do not have this issue:
488
489 [[BD3C-5 test, as above. There is no additional noise on the horizontal channels.>>image:BD3C_psd.png]]
490
491 = **Cleaning** =
492
493 When assembled, the nodes are water resistant but not submersible. They can handle a good spray and wipe-down. A stiff plastic brush is helpful to reach areas between the metal spikes on the bottom.
494
495 = **Weights (for shipping)** =
496
497 The weights of bags of nodes, as well as data harvesters and node chargers, are listed below:
498
499 1 bag + 6*IGU-16HR nodes: 18 kg
500
501 1*IGU-16HR data harvester: 21.5 kg
502
503 1*IGU-16HR charger: 26.3 kg
504
505 1*BD3C-5 charger (with and without 16 cables): 21 kg / 14.5 kg
506
507 1 case + 5*BD3C-5 nodes: 22 kg (aggregate battery weight <5kg, 168Wh)
508
509 1 case + 6*BD3C-5 nodes: 25 kg (aggregate battery weight >5kg, 168Wh)
510 )))
511
512 (% class="col-xs-12 col-sm-4" %)
513 (((
514 (% class="box" %)
515 (((
516 **Contents**
517
518 {{toc/}}
519
520
521 )))
522
523 (% class="box" %)
524 (((
525 = SmartSolo [[BD3C-5>>url:https://smartsolo.com/cp-4.html]] =
526
527 [[image:Smartsolo IGU BD3C 5 (2).jpg]]
528
529 [[image:smartsolo.jpg]]
530
531 |(% style="width:189px" %)**Frequency Band**|(% style="width:221px" %)5 Seconds to 150Hz
532 |(% style="width:189px" %)**Sensitivity**|(% style="width:221px" %)200 V/m/s
533 |(% style="width:189px" %)**Size (without spike)**|(% style="width:221px" %)Φ158 x160mm(H)
534 |(% style="width:189px" %)**Weight**|(% style="width:221px" %)2.8 kg
535 |(% style="width:189px" %)**Data Storage**|(% style="width:221px" %)64 Gb
536 |(% style="width:189px" %)**Battery**|(% style="width:221px" %)(((
537 Lithium-ion battery contained in equipment (168.84 Wh)
538
539 UN3481 PI967 S1
540 )))
541 )))
542
543 (% class="box" %)
544 (((
545 = SmartSolo [[IGU-16HR>>url:https://smartsolo.com/cp-3.html]]3C =
546
547 [[image:smartsolo node.jpg]]
548
549 [[image:smartsolo node 2.jpg]]
550
551 |(% style="width:187px" %)**Frequency Band**|(% style="width:224px" %)5 Hz to 1652Hz
552 |(% style="width:187px" %)**Sensitivity**|(% style="width:224px" %)76.7 V/m/s
553 |(% style="width:187px" %)**Size (with spike)**|(% style="width:224px" %)103mm(L) × 95mm(W) × 187mm(H)
554 |(% style="width:187px" %)**Weight**|(% style="width:224px" %)2.4 kg
555 |(% style="width:187px" %)**Data Storage**|(% style="width:224px" %)64 Gb
556 |(% style="width:187px" %)**Battery**|(% style="width:224px" %)(((
557 Lithium-ion battery contained in equipment (96.48 Wh)
558
559 UN3481 PI967 S2
560 )))
561 )))
562
563 (% class="box" %)
564 (((
565
566
567 = SmartSolo [[IGU-16>>url:https://smartsolo.com/cp-3.html]]1C =
568
569
570 [[image:Screenshot 2025-08-01 161027.png]]
571
572 |(% style="width:187px" %)**Frequency Band**|(% style="width:224px" %)5 Hz to 413Hz
573 |(% style="width:187px" %)**Sensitivity**|(% style="width:224px" %)80 V/m/s
574 |(% style="width:187px" %)**Size (without spike)**|(% style="width:224px" %)95mm(L) × 103mm(W) × 118mm(H)
575 |(% style="width:187px" %)**Weight**|(% style="width:224px" %)1.1 kg
576 |(% style="width:187px" %)**Data Storage**|(% style="width:224px" %)8 Gb
577 |(% style="width:187px" %)**Battery**|(% style="width:224px" %)(((
578 Lithium-ion battery contained in equipment (38.48 Wh)
579
580 UN3481 PI967 S2
581 )))
582 )))
583
584 (% class="box" %)
585 (((
586 = SmartSolo BD3C-16 Portable Battery Charger =
587
588 [[image:20250729_125049.jpg]]
589
590 |**Dimensions (LxHxW)**|558 x 357 x 300mm
591 |**Input rating**|100-210V - 50/60Hz
592 |**Power**|1000W
593 |**Weight**|14.5 kg
594 |**Weight with cables**|21 kg
595 )))
596
597 (% class="box" id="HSmartSoloBD3C-16PortableBatteryCharger" %)
598 (((
599 = SmartSolo IGU-16 Portable Data Harvester =
600
601 [[image:20250729_124747.jpg]]
602
603 |**Dimensions (LxHxW)**|625 x 500 x 366mm
604 |**Input rating**|100-210V - 50/60Hz
605 |**Power**|100W
606 |**Weight**|21.5 - 24 kg
607 |**Capacity**|16 nodes
608 |**Download Speed**|20MB/sec/slot
609 )))
610
611 (% class="box" %)
612 (((
613 = SmartSolo IGU-16 Portable Battery Charger =
614
615 [[image:20250729_124644.jpg]]
616
617 |**Dimensions (LxHxW)**|625 x 500 x 366 mm
618 |**Input rating**|100-210V - 50/60 Hz
619 |**Power**|640 W
620 |**Weight**|26.3 kg
621 |**Capacity**|16 nodes
622 )))
623
624 (% class="box" %)
625 (((
626 = SmartSolo BD3C-5 Carry Case =
627
628 [[image:20250729_124957.jpg]]
629
630
631 |**Dimensions (LxHxW)**|590 x 225 x 405 mm
632 |**Weight**|8.2 kg
633 |**Capacity**|6 nodes
634 )))
635
636 (% class="box" %)
637 (((
638 = SmartSolo IGU-16 3C Carry Bag =
639
640 [[image:20250729_124502.jpg]]
641
642 |**Dimensions (LxHxW)**|230 x 340 x 310mm
643 |**Weight**|(((
644 3.6kg (empty)
645
646 18.0kg (full)
647 )))
648 |**Capacity**|6 nodes
649 )))
650
651 (% class="box" %)
652 (((
653 = SmartSolo IGU-16 1C Carry Bag =
654
655 [[image:20250729_124558.jpg]]
656
657 |**Dimensions (LxHxW)**|225 x 200 x 550mm
658 |**Weight**|
659 |**Capacity**|8 nodes
660 )))
661 )))
662 )))