Changes for page SmartSolo Node Seismometers
Last modified by robert on 2026/01/19 17:43
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -23,11 +23,10 @@ 23 23 24 24 The nodes must be programmed in the SoloLite software prior to use. The screenshots below show our recommended parameters for the 5 Hz (16HR-3C) and 5 second (BDC3-5) nodes. 25 25 26 -[[IGU16 HR-3C programming screen set at 250hz. Ensure circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"data-xwiki-image-label="IGU-163C programming screensetat 250 Hz. Ensurethat the circledareasare set!"]]26 +[[IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]][[image:5S_node_programming.labels.png||alt="BD3C-5 programming screen"]][[image:5S_node_programming.labels.png]] 27 27 28 - [[BD3C-5 programming screen set at 250 hz. Ensurecircled areas are set!>>image:5S_node_programming.labels.png||alt="BD3C-5 programming screen set at 250hz. Ensure that the circled areas are set!"]]28 +BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set! 29 29 30 - 31 31 FIFO (first in, first out) data mode is safest as this will overwrite old data in case you forgot to clear the storage. At <= 250 hz you can fit 4++ months of data on these, shouldn't be an issue. 32 32 33 33 Note that the samplerate is instead given in sample spacing, in milliseconds. 4 ms = 250 Hz, 1 ms = 1000 Hz, 10 ms = 100 Hz, ad nauseam.