... ... @@ -23,10 +23,11 @@
23 23  
24 24  The nodes must be programmed in the SoloLite software prior to use. The screenshots below show our recommended parameters for the 5 Hz (16HR-3C) and 5 second (BDC3-5) nodes.
25 25  
26 -[[IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]][[image:5S_node_programming.labels.png||alt="BD3C-5 programming screen"]][[image:5S_node_programming.labels.png]]
26 +[[IGU16HR-3C programming screen set at 250 Hz. Ensure circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]]
27 27  
28 -BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!
28 +[[BD3C-5 programming screen set at 250 Hz. Ensure circled areas are set!>>image:5S_node_programming.labels.png||alt="BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!"]]
29 29  
30 +
30 30  FIFO (first in, first out) data mode is safest as this will overwrite old data in case you forgot to clear the storage. At <= 250 hz you can fit 4++ months of data on these, shouldn't be an issue.
31 31  
32 32  Note that the samplerate is instead given in sample spacing, in milliseconds. 4 ms = 250 Hz, 1 ms = 1000 Hz, 10 ms = 100 Hz, ad nauseam.
... ... @@ -68,6 +68,8 @@
68 68  
69 69  We have experienced interference from animals (foxes, dogs, goats) digging up and carrying nodes off for tens or hundreds of metres. It is helpful to minimise human and food smells (particularly on the rope handles) when working in areas where this is a risk. Or, wipe down affected nodes with 50-80% methylated spirits if extensive handling can not be avoided.
70 70  
72 +If you come to collect your node and it is missing~-~- LOOK FOR IT! It may not have gotten far. We have found dozens of nodes by spending 15 minutes looking for them.
73 +
71 71  = **Installation** =
72 72  
73 73  (% class="box infomessage" %)
... ... @@ -328,7 +328,7 @@
328 328  
329 329  See [[https:~~/~~/auspass.edu.au/xwiki/bin/view/Data/AusPass%20Data/#HSmartSoloNodeZPolaritybug>>https://auspass.edu.au/xwiki/bin/view/Data/AusPass%20Data/#HSmartSoloNodeZPolaritybug]] for discussion. If data is headed to AusPass, we prefer to invert the IGU-16HR channel data manually rather than in the SoloLite software or inverting the response metadata.
330 330  
331 -**The BD3C-5 data does not require a polarity inversion.**
334 +**The BD3C-5 data does not require any sort of polarity inversion.**
332 332  