... ... @@ -23,10 +23,11 @@
23 23  
24 24  The nodes must be programmed in the SoloLite software prior to use. The screenshots below show our recommended parameters for the 5 Hz (16HR-3C) and 5 second (BDC3-5) nodes.
25 25  
26 -[[IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]][[image:5S_node_programming.labels.png||alt="BD3C-5 programming screen"]][[image:5S_node_programming.labels.png]]
26 +[[IGU16HR-3C programming screen set at 250 hz. Ensure circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen" data-xwiki-image-label="IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!"]]
27 27  
28 -BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!
28 +[[BD3C-5 programming screen set at 250 hz. Ensure circled areas are set!>>image:5S_node_programming.labels.png||alt="BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!"]]
29 29  
30 +
30 30  FIFO (first in, first out) data mode is safest as this will overwrite old data in case you forgot to clear the storage. At <= 250 hz you can fit 4++ months of data on these, shouldn't be an issue.
31 31  
32 32  Note that the samplerate is instead given in sample spacing, in milliseconds. 4 ms = 250 Hz, 1 ms = 1000 Hz, 10 ms = 100 Hz, ad nauseam.