Changes for page SmartSolo Node Seismometers
Last modified by robert on 2026/01/19 17:43
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -126,10 +126,10 @@ 126 126 * **State of Charge (SoC) for Storage**: 127 127 ** Maintain a battery charge level of around 50-60% (i.e., ORANGE) for storage. 128 128 ** This charge level is recommended to prevent battery damage, and should be checked every six months. 129 -** Nodes should //__not be stored at full-charge (GREEN), or 0-charge (RED).__//129 +** Nodes should __not be stored at full-charge (GREEN), or 0-charge (RED).__ 130 130 ** Storage at 0-charge damages lithium batteries**.** 131 131 * **SoC for Transport:** 132 -** Charge levels for transport will be advised by the freighter. The required SoC will depend onvolume andtransportmethod (air, land, sea).132 +** Charge levels for transport will be advised by the freighter. The required SoC will usually fall between 0-60% depending on method (air, land, sea). 133 133 134 134 ((( 135 135 == 6. Data Sharing and Metadata Creation ==