... ... @@ -23,9 +23,9 @@
23 23  
24 24  The nodes must be programmed in the SoloLite software prior to use. The screenshots below show our recommended parameters for the 5 Hz (16HR-3C) and 5 second (BDC3-5) nodes.
25 25  
26 -[[IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]]
26 +[[IGU-16 3C programming screen set at 250 Hz. Ensure that the circled areas are set!>>image:5Hz_node_programming.labels.png||alt="IGU-16 3C programming screen"]][[image:5S_node_programming.labels.png||alt="BD3C-5 programming screen"]][[image:5S_node_programming.labels.png]]
27 27  
28 -[[BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!>>image:5S_node_programming.labels.png||alt="BD3C-5 programming screen"]]
28 +BD3C-5 programming screen set at 250 hz. Ensure that the circled areas are set!
29 29  
30 30  FIFO (first in, first out) data mode is safest as this will overwrite old data in case you forgot to clear the storage. At <= 250 hz you can fit 4++ months of data on these, shouldn't be an issue.
31 31  
... ... @@ -126,10 +126,10 @@
126 126  * **State of Charge (SoC) for Storage**:
127 127  ** Maintain a battery charge level of around 50-60% (i.e., ORANGE) for storage.
128 128  ** This charge level is recommended to prevent battery damage, and should be checked every six months.
129 -** Nodes should __not be stored at full-charge (GREEN), or 0-charge (RED).__
129 +** Nodes should //__not be stored at full-charge (GREEN), or 0-charge (RED).__//
130 130  ** Storage at 0-charge damages lithium batteries**.**
131 131  * **SoC for Transport:**
132 -** Charge levels for transport will be advised by the freighter. The required SoC will usually fall between 0-60% depending on method (air, land, sea).
132 +** Charge levels for transport will be advised by the freighter. The required SoC will depend on volume and transport method (air, land, sea).
133 133  
134 134  (((
135 135  == 6. Data Sharing and Metadata Creation ==
... ... @@ -415,21 +415,23 @@
415 415  
416 416  = **Cleaning** =
417 417  
418 -When still connected, the nodes are water resistant (don't submerge them!) and can handle a good spray / wipe-down. A strong, non-wire brush is helpful to reach areas between the metal spikes on the bottom.
418 +When assembled, the nodes are water resistant but not submersible. They can handle a good spray and wipe-down. A strong, non-wire brush is helpful to reach areas between the metal spikes on the bottom.
419 419  
420 420  = **Weights (for shipping)** =
421 421  
422 422  The weights of bags of nodes, as well as data harvesters and node chargers, are listed below:
423 423  
424 -1 bag + 6 SP (IGU-16HR) nodes: 18 kg
424 +1 bag + 6*IGU-16HR nodes: 18 kg
425 425  
426 -1 SP (IGU-16HR) data harvester: 21.5 kg
426 +1*IGU-16HR data harvester: 21.5 kg
427 427  
428 -1 SP (IGU-16HR) charger: 26.3 kg
428 +1*IGU-16HR charger: 26.3 kg
429 429  
430 -1 BB (BD3C-5) charger/data harvester (with and without 16 cables): 21 kg / 14.5 kg
430 +1*BD3C-5 charger (with and without 16 cables): 21 kg / 14.5 kg
431 431  
432 -1 case + 5 BB (BD3C-5) nodes and 6 BB nodes: 22 kg / 25 kg
432 +1 case + 5*BD3C-5 nodes: 22 kg (aggregate battery weight <5kg, 168Wh)
433 +
434 +1 case + 6*BD3C-5 nodes: 25 kg (aggregate battery weight >5kg, 168Wh)