... ... @@ -221,14 +221,15 @@
221 221  * Go to the “Tool” menu and select “export seismic data”.
222 222  * Tailor other parameters to personal preference and ensure "Sample Interval" matches the setting used during node reset.
223 223  * Click “prepare” followed by “run” to start reformatting. Monitor this process in the small panel at the bottom left.
224 +)))
225 +
224 224  * (% class="box warningmessage" %)
225 225  (((
226 -* **Ensure to export data as "COUNTS", not "mV".**
227 -
228 -
229 -
230 -* **Set "Remove Gain" to the same decibel gain as during programming (ANU usually sets this to 18 db, but double-check).**
228 +Ensure to export data as "COUNTS", not "mV".
231 231  )))
230 +* (% class="box warningmessage" %)
231 +(((
232 +Set "Remove Gain" to the same decibel gain as during programming (ANU usually sets this to 18 db, but double-check).
232 232  )))
233 233  
234 234  ==== **Handling Nodes During Download** ====
... ... @@ -245,11 +245,13 @@
245 245  
246 246  * (% class="box warningmessage" %)
247 247  (((
248 -* **Use fast external hard drives to avoid limitations in data harvesting.**
249 -
250 -* **Recommended specifications: USB-C, USB 3.0, and 4+ Tb of space.**
249 +Use fast external hard drives to avoid limitations in data harvesting.
251 251  )))
251 +* (% class="box warningmessage" %)
252 +(((
253 +Recommended specifications: USB-C, USB 3.0, and 4+ Tb of space.
252 252  )))
255 +)))
253 253  1. (((
254 254  **Metadata and Time Settings**:
255 255  
... ... @@ -259,9 +259,15 @@
259 259  1. (((
260 260  **Finalizing the Download**:
261 261  
262 -* After downloading, mark the //"D"// box on your temporary labels to indicate completion.
265 +* After downloading, mark the "D" box on your temporary labels to indicate completion.
263 263  )))
264 264  
268 +
269 +
270 +
271 +
272 +
273 +
265 265  = SmartSolo IGU 16HR 3C (5 Hz) Short Period Node =
266 266  