... ... @@ -4,11 +4,8 @@
4 4  (((
5 5  = **Node Setup** =
6 6  
7 -Short term (~~30 day) battery powered nodes. The default gain should be 24 db for the 16HR-3C and 6 db for the BD3C-5
7 +Short term (~~30 day) battery powered nodes
8 8  
9 -
10 -~*~**this needs to be reformatted
11 -
12 12  == **Pre-Fieldwork Preparation** ==
13 13  
14 14  (% class="box infomessage" %)
... ... @@ -15,9 +15,7 @@
15 15  (((
16 16  (% class="box warningmessage" %)
17 17  (((
18 -**INVEST IN FAST EXTERNAL HARD DRIVES – DO NOT LET THIS BE THE LIMITATION OF DATA HARVESTING**
19 -
20 -**Plan on 1 Tb to store ALL data for 50 nodes @ 250 Hz and a single ~~30 day occupation**
15 +**INVEST IN FAST EXTERNAL HARD DRIVES – DO NOT LET THIS BE THE LIMITATION OF DATA HARVESTING **
21 21  )))
22 22  )))
23 23  
... ... @@ -66,13 +66,16 @@
66 66  * **Pre-Deployment Charging**:
67 67  ** Although the nodes hold their charge well, it's beneficial to give them a "top up" charge right before deployment.
68 68  
64 +
69 69  * **Operational Duration**:
70 70  ** When recording at 250 Hz, with GPS on and Bluetooth disabled, the instruments are expected to last about 30 days per charge cycle.
71 71  
68 +
72 72  * **Post-Retrieval Charging**:
73 73  ** Do not store the instruments with completely drained batteries.
74 74  ** After retrieval, charge the instruments to about 50-60% (indicated as "orange" level) for optimal battery health.
75 75  
73 +
76 76  * **Storage and Shipping Charge Level**:
77 77  ** Maintain a battery charge level of around 50-60% for both storage and shipping purposes.
78 78  ** This charge level is recommended to prevent battery damage and is safe for transportation.