Sima Mousavi 6.5 1 = **What is iDAS?** =
2
3 **iDAS** stands for **Integrated Distributed Acoustic Sensing**. This advanced system utilizes the principle of Rayleigh backscattering in optical fibers to measure strain rates along the fiber's length. It's a cutting-edge technology predominantly used in fields like geophysics, where it's crucial for monitoring seismic activities, as well as in industrial applications like pipeline monitoring and security systems.
4
Sima Mousavi 8.1 5 === **Safety first!** ===
Sima Mousavi 6.5 6
Sima Mousavi 8.1 7 === ===
8
Sima Mousavi 6.5 9 (% class="box warningmessage" %)
10 The iDAS uses a Class 1 laser, which is generally considered safe for most applications. However, direct exposure to the eye should be avoided to prevent any risk of eye injury.
11 \\Cables and fibers may create tripping hazards. Ensure that the routing of cables is done in a way that minimizes the risk of accidents.
12
13 ----
14
Sima Mousavi 7.1 15 === **System Components** ===
16
17
Sima Mousavi 8.2 18 The iDAS system consists of the following components:
Sima Mousavi 7.1 19
Sima Mousavi 8.2 20 1. **Interrogator unit:** This unit houses the laser, detector, and electronics for processing the backscattered light signal. It records various settings such as the optical system configurations, spatial sampling rates, and the duration of data acquisition. More than one interrogator may be used in an experiment. Each unit should have a unique identifier and is described in its respective metadata block.
21 1. **Optical fiber:** The optical fiber is the sensor that is used to measure strain rate.
22 1. **Data acquisition software**: This software is used to control the interrogator unit and to acquire and process the data. It records various settings such as the optical system configurations, spatial sampling rates, and the duration of data acquisition.
Sima Mousavi 7.1 23
Sima Mousavi 8.2 24 ----
Sima Mousavi 7.1 25
Sima Mousavi 6.3 26 = **DAS Data Collection System** =
Sima Mousavi 5.1 27
Sima Mousavi 8.2 28 ==== **1. Surface Installation** ====
Sima Mousavi 6.3 29
30 * **Cable Arrangement:** The optical fiber cable can be a single long continuous cable or multiple cables with connectors, used across different installation environments.
31 * **Cable Types:** Cables may vary in construction types based on the installation environment.
32 * **Geographical Metadata:** Parameters related to geographical location are specified for easy searchability in databases.
33
Sima Mousavi 8.2 34 ==== **2. Installation Environments** ====
Sima Mousavi 6.3 35
36 * **Shallow Burial:** The cable can be shallowly buried.
37 * **In-Conduit Installation:** The cable can also be installed within a conduit.
38 * **Water Installation:** Suitable for underwater environments.
39
Sima Mousavi 8.2 40 ==== **3. Channels** ====
Sima Mousavi 6.3 41
42 * **Sensor Distribution:** Channels represent sensors distributed along the fiber array at fixed intervals.
43 * **Physical Location Determination:** The physical locations of these channels are ascertainable through methods like tap tests and GPS tracking.
44 * **Metadata for Channels:** Each channel or group of channels has its own metadata block, detailing specifics like location and properties.
45
Sima Mousavi 6.5 46 [[image:1706489800616-982.png]]
47
48
49
Sima Mousavi 8.1 50 [[image:1706488566274-805.png]]
Sima Mousavi 6.5 51
Sima Mousavi 6.4 52 ----
Sima Mousavi 6.3 53
Sima Mousavi 6.4 54 = **DAS Metadata Schema** =
55
56 ==== **1. Interrogator** ====
57
58 * **Required:** Interrogator ID, manufacturer, model.
59 * **Optional:** Serial number, firmware version.
60
61 ==== **2. Acquisition Parameters** ====
62
63 * **Required:** Interrogator ID, acquisition ID, unit of measure, acquisition start and end time, acquisition sample rate, gauge length, number of channels, spatial sampling interval.
64 * **Optional:** Pulse rate, pulse width.
65
66 ==== **3. Overview** ====
67
68 * **Required:** Network code, location, start date, end date, point of contact.
69 * **Optional:** Digital Object Identifier (DOI), purpose of data collection, funding agency.
70
71 ==== **4. Cable Information** ====
72
73 * **Required:** Cable ID, cable bounding box.
74 * **Optional:** Cable owner, cable installation date, cable characteristics.
75
76 ==== **5. Fiber Information ('Native format')** ====
77
78 * **Required:** Cable ID, fiber ID, fiber geometry, fiber mode, fiber refraction index.
79 * **Optional:** Fiber optical length, fiber start and end location, fiber one-way attenuation, fiber winding angle.
80
81 ==== **6. Channel Group** ====
82
83 * **Required:** Interrogator ID, acquisition ID, channel group ID, cable ID, fiber ID, coordinate generation date.
84 * **Optional:** Location method, coordinate reference frame, first/last usable channel ID, uncertainty of various coordinates (x, y, latitude, longitude, elevation, depth, strike, dip).
85
86 ==== **7. Channel** ====
87
88 * **Required:** Channel group ID, channel ID, coordinates (x, y, latitude, longitude).
89 * **Optional:** Elevation above sea level, depth below surface, distance along fiber, strike, dip.
90
Sima Mousavi 5.1 91 ----
92
Sima Mousavi 6.3 93 === ===
Sima Mousavi 5.1 94
Sima Mousavi 6.3 95 = **Pre-processing Steps for DAS Data (2.5 months, 90 TB)** =
96
97 1. (((
98 **Create a Deployment Log**
99
100 * Document key details like photonics settings, duration of data collection, instances of hard disk failure, power outages, etc.
101 )))
102 1. (((
103 **Generate Geographic Coordinates**
104
105 * Assign latitude and longitude coordinates to each DAS channel for spatial referencing.
106 )))
107 1. (((
108 **Data Conversion**
109
110 * Convert the raw DAS time series data from counts to strain rate, which is more meaningful for seismic analysis.
111 )))
112 1. (((
113 **Data Decimation**
114
115 * Reduce the sampling rate from 1000 Hz to 100 Hz. This step is essential to manage data volume and focus on frequencies of interest for seismic activities.
116 )))
117 1. (((
118 **Extract Specific Time Series**
119
120 * Isolate time series data of special interest, such as those related to earthquake events, for detailed analysis.
121 )))
122 1. (((
123 **Data Hosting**
124
125 * Store the processed data on Research School of Earth Sciences (RSES) clusters and the National Computational Infrastructure (NCI) for accessibility and further analysis.
126 )))
127 1. (((
128 **Extract Data and Instrument Response from Seismic Nodes**
129
130 * Retrieve both the raw data and the corresponding instrument response characteristics for accurate interpretation of the seismic signals.
131 )))
132 1. (((
133 **Long-term Archiving and Data Sharing**
134
135 * Ensure that the data is archived in a format and location that supports long-term preservation and accessibility, such as the AuScope Distributed Acoustic Sensing (DAS) Collection.
136 )))
137 1. (((
138 **Data Access and Record Keeping**
139
140 * Maintain records and access paths (e.g., NCI local file path) for the data, including metadata and other associated records, to facilitate future research and data sharing.
141 )))
142
143 === ===
144
Sima Mousavi 9.15 145 ----
Sima Mousavi 6.3 146
Sima Mousavi 9.16 147 == **iDAS Software Overview** ==
Sima Mousavi 9.11 148
Sima Mousavi 9.16 149 === **Operating System Compatibility:** ===
Sima Mousavi 9.15 150
151 * (((
152 The iDAS system software is designed to run on the Windows operating system, with the current version compatible with Windows 7.
153 )))
154
Sima Mousavi 9.11 155 === **Accessing the System:** ===
156
157 * **User Login:** The system is set up to log in automatically as "User", without requiring any credentials.
158 * **Software Shortcut:** A shortcut to access the iDAS software is conveniently placed on the desktop for easy access.
159
Sima Mousavi 9.16 160 === **Data Acquisition Controls** ===
Sima Mousavi 6.3 161
Sima Mousavi 9.16 162 * (((
163 **Start / Stop:** Easily control the start and stop of data acquisition using the current settings.
164 )))
Sima Mousavi 9.15 165 * **Calibrate Reference:** Recalibrate reference levels as needed.
166 * **Exit:** Safely exit the iDAS software.
167
Sima Mousavi 9.16 168 === **Configuration iDAS Setup Wizard:** Provides a guided, step-by-step process for setting up the iDAS system. ===
Sima Mousavi 9.15 169
170 * **iDAS Configuration:** Access the iDAS configuration window for advanced settings.
171 * **Trigger Mode:** Opens the trigger configuration window for specific acquisition needs.
172
Sima Mousavi 9.16 173 === **Operating Modes ** ===
Sima Mousavi 9.15 174
175 * **Available Modes:** Acoustic, Waterfall, Stacking, Security, and OTDR.
176 * **Functionality:** Each mode presents the acoustic data from iDAS in unique formats.
177 * **Detailed Information:** See pages 19-23 of the Standard iDAS manual for comprehensive details on each mode.
178
Sima Mousavi 6.3 179 === ===
180
181 === ===
182
183 ----
184
185 = **Silixa iDAS Setup Manual for Reoccupying/Duplicating Experiment** =
186
187 = **General Information** =
188
Sima Mousavi 5.1 189 * **Experiment Name:** ~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~__
190 * **Local Date:** ~_~_~_~_/~_~__**/202**
191 * **Local Time:** ~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~__
192 * **Personnel:** ~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_
193 * **Site Condition Notes:**
194 ** [Space for detailed notes about the site condition, any upkeep performed, and changes made since the last visit.]
195
196 ----
197
198 ==== **Setup Procedure** ====
199
200 1. (((
201 **Open iDAS Box**
202
203 * Serial Number: ~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~__
204 * IP Address: ~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_~_
205 )))
206 1. (((
207 **Check Laser Lock**
208
209 * Ensure the lever on the LASER/RIGHT side of the iDAS is in the locked position (LEVER UP).
210 )))
211 1. (((
212 **Placement of iDAS**
213
214 * Position the iDAS on a stable surface.
215 * Ensure clear air flow around the fan vents.
216 )))
217 1. (((
218 **UPS Setup** (If applicable)
219
220 * Connect the UPS to a power source.
221 * Plug the NAS into the UPS using the special cable with a rectangular connector.
222 * Connect the monitor to the UPS.
223 )))
224 1. (((
225 **Peripheral Connections**
226
227 * Connect a computer monitor, mouse, and keyboard to the iDAS.
228 )))
229 1. (((
230 **GPS Antenna Connection**
231
232 * Connect the GPS antenna to the left-side socket on the front of the iDAS using a special adapter.
233 )))
234 1. (((
235 **Fibre Optic Connection**
236
237 * Clean both the fibre optic cable and the socket with a 1-click cleaner before connection.
238 * Use the guide/tip for cleaning the fibre and clean the socket without it.
239 )))
240 1. (((
241 **Powering the iDAS**
242
243 * Connect the iDAS to a power socket or UPS.
244 * Note: The iDAS turns on automatically when connected to power. Use the power button at the bottom left if it does not.
245 )))
246 1. (((
247 **Network Drive Mapping**
248
249 * Open File Explorer and select 'This PC' from the left pane.
250 * On the File Explorer ribbon, select 'More' > 'Map network drive'.
251 * Choose a drive letter from the 'Drive' list.
252 )))
253 1. (((
254 **Open iDAS Software**
255
256 * Use the link/shortcut on the Desktop.
257 )))
258 1. (((
259 **GPS Satellite Confirmation**
260
261 * Check for a green light in the top row of the software interface indicating GPS satellite detection.
262 )))
263 1. (((
264 **Lever/Interlock Adjustment**
265
266 * Lower the lever/interlock (as mentioned in step 2) before starting data acquisition.
267 )))
268 1. (((
269 **Configuration File Duplication**
270
271 * Use 'current.cfg' for duplicating previous setup.
272 * The file is located at C:/Silixa/Data/current.cfg.
273 * Copy this file to the desktop and save it with a new filename.
274 )))
275 1. (((
276 **iDAS Configuration Settings**
277
278 * Navigate to the 'Configuration' tab and click 'iDAS configuration'.
279 * Check all values under 'system'.
280 * Under 'application', enable error messages.
281 * In 'saving', select 'acoustic saving', choose 'tdms', and set the file save location and identifier.
282 * Leave default parameters in 'measure'.
283 )))
284 1. (((
285 **Mode Selection**
286
287 * Toggle between Acoustic mode, Waterfall mode, and OTDR mode to check data recording.
288 )))
289 1. (((
290 **Photograph OTDR Mode**
291
292 * Document the OTDR mode settings with a photograph.
293 )))
294 1. (((
295 **Photograph Waterfall Mode**
296
297 * Similarly, take a picture of the Waterfall mode settings.
298 )))
299 1. (((
300 **Configuration File Saving**
301
302 * Save the configuration file in .cfg format in a separate location without overwriting the existing one.
303 )))
304 1. (((
305 **Software Check**
306
307 * Ensure QFinder, Teamviewer, and Microsoft Remote Desktop are installed.
308 )))
309 1. (((
310 **Remote Access Setup**
311
312 * Open Teamviewer or Microsoft Remote Desktop.
313 * For Teamviewer, photograph the ID and password and ensure it remains open.
314 * For Microsoft Remote Desktop, record the IP address and test remote access with a separate laptop.
315 )))
316 1. (((
317 **Final Mode Setting**
318
319 * Switch to OTDR mode for ongoing operation (avoids excessive memory usage of Waterfall mode).
320 )))
321
322 ----
323
324 ==== **Important Reminders** ====
325
326 * Always ensure that the iDAS is connected to a suitably earthed power socket or UPS for safety.
327 * Regularly check the condition of all cables and connectors.
328 * Before leaving the setup, confirm that data saving is active if required.
329 * Keep a log of all actions taken during the setup for future reference and troubleshooting.
Sima Mousavi 6.1 330
331
332
333
334
Sima Mousavi 6.2 335
336
Sima Mousavi 6.1 337
Sima Mousavi 6.2 338 [[Caption>>image:1706488806394-168.png]]
Sima Mousavi 6.1 339
340
341 [[image:https://photos.fife.usercontent.google.com/pw/ABLVV86phmYsze0lp1yzL-gs6stE23rmgKbOxcpCRhFmK9s25zf5UHBkgbFcOg=w1440-h2556-s-no-gm?authuser=0]]
342
343
344
Sima Mousavi 6.2 345 [[How to clean the optical connector>>image:1706488909085-143.png||alt="How to clean the optical connector"]]
Sima Mousavi 6.1 346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
Sima Mousavi 6.3 362