... ... @@ -254,20 +254,49 @@
254 254  
255 255  
256 256  
257 +== **Integrating GPS Antenna for Accurate Timing** ==
257 257  
259 +=== **Installation and Connection Process** ===
258 258  
261 +**Connecting the GPS Antenna:**
259 259  
263 +* Connect one end of a coaxial cable with SMB female connectors to the GPS antenna.
264 +* If you plan to mount the antenna, do so before connecting the cable.
265 +* Connect the other end of the cable to the 'GPS Ant.' port located on the output panel at the front of the iDAS unit.
260 260  
267 +=== **Monitoring GPS Status and Time** ===
261 261  
269 +* **GPS Information Display:** The GPS status and time are displayed on the Tool Bar of the iDAS software.
270 +* **Status Indicators:**
271 +** **Red:** Antenna is not connected.
272 +** **Amber:** Antenna connected, but self-survey not completed.
273 +** **Green:** Antenna connected, self-survey completed, and timing is accurate.
262 262  
275 +[[image:1706509432197-316.png]]
263 263  
264 264  
265 265  
279 +**Primary Configuration Settings in iDAS**
266 266  
281 +These settings are integral to the operation of the iDAS system, significantly influencing aspects like the rate of data streaming. They encompass:
267 267  
283 +1. **Sampling Frequency:** Determines how often data is captured.
284 +1. **Spatial Resolution:** Defines the level of detail in spatial data.
285 +1. **Fiber Length:** Impacts the range over which data is collected.
268 268  
287 +In configuring these parameters, a balance must be struck. This involves aligning the system's physical capabilities, such as fiber length and spatial resolution, with the practical limits of data collection, including the volume of data generated. Adjusting these settings correctly is crucial for optimizing data quality while managing the data volume effectively. This balance ensures that the system meets the specific needs of each application without overwhelming storage or processing capabilities.
269 269  
270 270  
290 +[[image:1706510182174-365.png]]
291 +
292 +
293 +
294 +[[image:1706510243501-784.png]]
295 +
296 +
297 +
298 +
299 +
271 271  === [[image:1706506304897-924.png]] ===
272 272  
273 273  [[image:1706506500585-231.png]]
1706509432197-316.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.sima
Size
... ... @@ -1,0 +1,1 @@
1 +2.0 MB
Content
1706510182174-365.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.sima
Size
... ... @@ -1,0 +1,1 @@
1 +1.2 MB
Content
1706510243501-784.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.sima
Size
... ... @@ -1,0 +1,1 @@
1 +1.6 MB
Content