Wiki source code of ANU Seismic Data Loggers

Version 77.1 by Jack Dent on 2026/09/22 13:26

Hide last authors
robert 42.1 1 (% class="row" %)
robert 40.1 2 (((
robert 42.1 3 (% class="col-xs-12 col-sm-8" %)
4 (((
5 (% class="jumbotron" %)
6 (((
7 (% class="container" %)
8 (((
robert 38.1 9 ANU has designed and built three generations of seismic data recorders, two of which (the small yellow TerraSAWR and the larger beige LPR-200) are still in use today. Despite their contrasting appearance, they both use the same software, data cards, and have the same instrument response.
robert 40.1 10 )))
robert 42.1 11 )))
robert 1.1 12
robert 46.2 13 = Power Considerations =
14
Jack Dent 74.1 15 Both the LPR-200 (or "Low Power Recorder" 200) and TerraSAWR are designed to use as little power as possible.
robert 46.2 16
santi 54.1 17 At 100 Hz and with a GPS cable connected these loggers draw about 220 mW of power once the screen is off (higher sample rates draw more power but only marginally, < 5 mW). Adding a sensor (e.g. a Trillium Compact 120) increases this to approximately 400 mW, or 0.4 volt-amps. So, in theory 7 Ah battery should last about 10 days without a solar panel, but in practice it seems to be a bit closer to 8 which may be due to variability in power drain while in getting GPS locks.
robert 46.2 18
Jack Dent 74.1 19 For very sunny environments (latitudes < 30) a 20 Volt, 10 Watt solar panel should have no issue keeping these loggers alive over the summer months, and assuming unobstructed skies, should also be fine over winter. **However** when possible a 20 Watt panel works best, especially as they usually cost only $10-20 more these days.
robert 47.1 20
Jack Dent 74.1 21 Measurements of a 20 Watt panel in the Canberra (Aus) area show that in full sun, roughly 0.75 amps are put out. As the LPR-200 and TSAWR loggers step down the solar voltage to 12v, a 20 Watt solar panel realistically provides 9 Watts to a seismic station. Theoretically, 1 hour and 9 minutes of sun per day should keep the battery charge stable. However, in practice, 1 hour and 30 minutes of full sunlight for a 20 Watt solar panel is required to keep these stations at a stable charge (this was tested over a 5 day period under Canberra sun in spring)
22
Jack Dent 75.1 23 **Note**: It is important to understand 2 main limitations of the solar component of these seismic stations. First, the voltage output of most solar panels will drop significantly if even a small portion of the panel is shaded. This is why 'full' sun is a requirement. Second, 1 hour and 30 minutes of full sun per day will not reliably be achieved each day. Thus, the sunniest possible location for a seismic station is always ideal.
Jack Dent 74.1 24
robert 71.2 25 40 Watt panels can also be used, and may be needed for extreme climates, or areas without a full sky view. 60 Watt panels have also been known to work, but this is now reaching the the limit of what the loggers can reasonably handle for extended periods and are not recommended for long term deploys.
26
robert 47.1 27 (% class="box infomessage" %)
28 (((
29 Power issues are easy and cheap to solve relative to the cost of your experiment, don't skimp!
30 )))
31
Jack Dent 52.1 32 (% class="wikigeneratedid" %)
santi 54.1 33 In the case of an LPR, there is a large compartment for housing an internal battery, able to accommodate anything from a 10-30Ah battery. To use a standard lead acid battery with a positive and negative terminal, a 6 pin adaptor must be used. This ensures the voltage from the external power port (pins A and C) connect to the battery and ensure the system actually recharges. (See [[Peripheral Equipment>>doc:Instrumentation.Peripheral Equipment.WebHome]] for a more comprehensive overview of this kind of setup)
Jack Dent 52.1 34
Jack Dent 76.1 35 === Recharging a logger battery ===
36
37 If a logger stops receiving solar power, the station will drain and the battery may become unusable. Upon servicing a station in the field, the voltage of the 12v lead-acid battery should be assessed while not powering anything. If the voltage is below 10.5v, it is likely not suitable for use in the field and should be replaced with a functional battery. If the voltage is below 12.5v, it is a good idea to replace it with a fully charged 12v battery (though if this is not possible, it is ok to leave this battery so long as the station was alive on arrival.
38
Jack Dent 77.1 39 If a 12v lead-acid battery is collected from the field and is still above 10.5v, recharge it fully using a 12v battery charger. These are easily acquired at auto or electronics store. Recharging a 12v lead-acid battery with a capacity of about 10Ah should take a couple of hours (~~2.5hrs).
Jack Dent 76.1 40
robert 65.1 41 = GPS Considerations =
42
robert 67.1 43 GPS is required for the data to have accurate timestamps. A standard 3-5V 1575.42 Mhz coaxial "dongle" antenna works fine and can be found for relatively cheap (e.g. [[https:~~/~~/www.elecbee.com/en-3555-gps-antenna-bnc-male-for-garmin-gps-120120xl125-sounder-with-cable-2m) >>https://www.elecbee.com/en-3555-gps-antenna-bnc-male-for-garmin-gps-120120xl125-sounder-with-cable-2m]]. Anything that locates satellites and gives you a GPS-synced time works!
robert 65.1 44
45 The TerraSAWR has a built-in GPS but this doesn't work as well, especially if the logger is (wisely) buried. **The LPR does NOT have a built-in GPS antenna, so an external antenna is mandatory.**
46
47 In a pinch, a severed or broken antenna can be mended back together relatively easily by non-experts. Even stripping the wire and twisting it back together by hand on site is possible!
48
robert 15.1 49 = Data Card Formatting and Information =
50
robert 42.1 51 Both the TerraSAWR and LPR-200 require SD Cards to be formatted in FAT32 filesystem. For 64Gb cards it can be difficult to format in FAT32, but [[software >>http://auspass.edu.au/field/fat32cardformatter.exe]]is available. ANU recommend SanDisk Extreme 150 mb/s cards in either 32 or 64Gb size. We strongly discourage using cards larger than 64Gb, and in general smaller cards are less likely to fail. We have also found that "adapter" cards (e.g. SD to microSD) are prone to having write issues and **strongly** advise against them.
robert 15.1 52
robert 42.1 53 The loggers can be "pre-programmed" with information (e.g. site name, sampling rate, etc) or they can be programmed in the field using the buttons on the logger. To pre-program the cards you simply edit a text file (named "[[ANUSRSetup.txt>>http://auspass.edu.au/field/ANUSRSetup.txt]]" for the LPRs, or "[[tSAWRSetup.txt>>http://auspass.edu.au/field/tSAWRSetup.txt]]" for the TerraSAWRs) and place it in the root directory on the SDCard. When the logger boots up, it will parse and load this information.
robert 24.1 54
robert 66.1 55 (% class="wikigeneratedid" id="HTheformatforANUSRSetup.txt2FLPR200swillbeasinglelineoftextthatlookslikethis:" %)
56 The format for ANUSRSetup.txt / LPR200s will be a single line of text that looks like this:
robert 24.1 57
robert 25.1 58 {{{XXX195G0100010034864 2 }}}
robert 24.1 59
robert 46.2 60 //corresponding to//
robert 24.1 61
robert 46.2 62 site **XXX19**,
63 network **5G**,
64 sample rate **0100**, (i.e. 100 Hz)
65 "gps interval" **01** (once per hour~-~- don't change),
66 start mode (always **0**),
67 stop mode (always **0**),
68 seismometer type (**3** = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
69 and seismometer serial number (**4864**).
70
Jack Dent 29.1 71 (% class="box warningmessage" %)
72 (((
robert 72.1 73 NOTE: the 2 at the very end is for "RECORD ON RESTART". The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume. (# of blank spaces before this doesn't matter)
Jack Dent 29.1 74 )))
robert 24.1 75
robert 66.1 76 (% class="wikigeneratedid" id="HTheformatforTSAWRloggersisshorter:" %)
77 The format for TSAWR loggers is shorter:
robert 24.1 78
Jack Dent 31.2 79 (% class="box errormessage" %)
80 (((
81 **NOTE: The format is shorter as they don't have an option to set the "GPS sync". The GPS sync number should be omitted otherwise the formatting will not work!**
82 )))
robert 24.1 83
Jack Dent 31.2 84 SITE15G02500039999 2
85
Jack Dent 33.1 86 corresponding to site SITE1, 
Jack Dent 31.2 87
88 network 5G
Jack Dent 29.1 89 sample rate 0250, (i.e. 250 Hz)
90 start mode (always 0),
91 stop mode (always 0),
Jack Dent 31.2 92 seismometer type (3 = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
Jack Dent 29.1 93 and seismometer serial number (9999).
robert 24.1 94
Jack Dent 29.1 95
robert 66.1 96 (% class="wikigeneratedid" id="HTheformattingprocessusingthelogger:" %)
97 The formatting process using the logger:
Jack Dent 29.1 98
Jack Dent 30.1 99 The process for formatting an SD card within the logger is straightforward. Navigate to the "SD INFORMATION" screen and press ERASE SD CARD. This process may take up to a minute. This will result in erasing all files from the card. Upon starting recording, a new 'seed' will be written containing all the information that the logger has been set with FINISH THIS SECTION
Jack Dent 29.1 100
robert 42.1 101 [[image:original_0747763c-e3dd-4667-b897-833f12c8e0b8_20241219_130604.jpg||height="240" width="298"]]
Jack Dent 29.1 102
robert 24.1 103 Here is a general guideline for how much data you can expect to fit on a card:
104
105 {{{ 64Gb card: 245 days @ 250hz or 610 days @ 100hz
Jack Dent 31.1 106 32Gb card: 122 days @ 250hz or 305 days @ 100hz
107 }}}
robert 24.1 108
robert 46.2 109 == Default Settings ==
robert 24.1 110
robert 46.2 111 As of 2025, the default settings for both LPR (v. 2.7) and TSWAR (v 3.6a) loggers are below. If you're using a broadband instrument, you are essentially good to go without having to program the cards.
112
113 * XX.ANUSR network and station name
114 * 100 Hz Sample Rate
santi 54.1 115 * 40 Vpp (or +/- 20 V) gain / Trillium Compact seismometer version
robert 46.2 116 * Record on Restart enabled
117
118 Note that if a user sets the gain incorrectly, this can be fixed later (assuming nothing clipped) by multiplying or dividing by factors of 2. The gain setting can be looked up from the logfile, else you may have to guess from a PSD or other method.
119
Jack Dent 31.1 120 = Logger Menus Overview (and setup) =
robert 24.1 121
Jack Dent 36.1 122 The same menus are used in both the TSAWR and LPR-200 loggers.
Jack Dent 31.4 123
Jack Dent 31.1 124 == System Information ==
125
126 This menu provides an overview of the status of the station. Under 'Initialisation Parameters', the status of systems that are initialised upon powering on the logger can be checked.
127
128 This menu also displays the firmware version, battery, external, and solar voltages, and the temperature of the system.
129
robert 59.2 130 * Check that all //Initialisation Parameters// are marked as successful.
131 * Check that solar voltage is above 10 V in the software, or preferrably physically check that the battery's voltage is increasing via a DMM.
Jack Dent 31.1 132
133 == Live Seismometer Data ==
134
135 This screen displays real time seismometer data for all 3 axes of the seismometer. Initially, a plot of raw data from all 3 channels is shown. By pressing enter, a high pass filter can be applied showing a more useful plot of all channels. The up and down arrows can be used to change between a view of all channels, individual channels, and all channels on different X axes.
136
robert 44.1 137 * Check that all 3 channels are present and are producing a signal.
138 * Stomp on the ground near where the sensor is installed to check that a signal is being picked up by all 3 channels.
139 * Common representations of poor signals include channels sloping smoothly from high values to zero, or sharp corners in the plots.
Jack Dent 31.1 140
Jack Dent 33.1 141 == GPS Data ==
Jack Dent 31.3 142
143 This menu displays the status of the stations' GPS connection. The screen lists; UTC time, UTC date, latitude, longitude, altitude, number of satellite connections, and SNR.
144
robert 59.2 145 * Check that the station is connected to satellites. 3 or more should be perfectly adequate to keep time.
Jack Dent 31.3 146
147 == SD Information ==
148
149 The menu lists if the SD card has been initialised, as well as the capacity and space free on the memory card. See "Data card formatting" (above) to see how to format and initialise the inserted SD card.
150
Jack Dent 31.2 151 == System Configuration ==
152
153 This menu is used to set the stations identifiers and parameters.
154
155 First, the sampling rate can be chosen from a set of options (1 Hz, 10 Hz, 25 Hz, 40 Hz, 50 Hz, 100 Hz, 250 Hz, 1000 Hz).
156
157 Next, the station identifier (up to 5 characters) and the network code (2 characters) can be set.
158
159 Record start and stop mode should be left to 'On Request' mode.
160
161 The seismometer model and serial number (up to 10 characters) can be set. Seismometer model options include:
162
santi 54.1 163 * Trillium Compact (same for 20s and 120s models)
robert 43.1 164 * CMG - 3ESP
165 * Guralp 40T
166 * LE-3D Lite
167 * Mark L4C
168 * Mark L4
Jack Dent 31.2 169
170 Additionally, the following settings can be used in place of the seismometer model types: +/- 20V, +/- 10V, +/- 5V
171
Jack Dent 31.3 172 The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume.
Jack Dent 31.2 173
robert 43.1 174 * Set the sampling rate, station identifier, and network code. Set the appropriate seismometer type. Most importantly, ensure the record on restart setting is checked.
Jack Dent 31.2 175
Jack Dent 31.3 176 (% class="box errormessage" %)
Jack Dent 31.2 177 (((
robert 60.1 178 Again, ensure the 'RECORD ON RESTART' option is marked with a cross (the default setting). This ensures that the logger will record any time it receives enough power!
Jack Dent 31.2 179 )))
180
Jack Dent 34.1 181 (% class="wikigeneratedid" %)
robert 42.1 182 [[image:original_2b83fcf5-ee5a-4375-bab1-2d738e4ffe8c_20241219_130646.jpg||height="234" width="289"]]
Jack Dent 34.1 183
Jack Dent 31.3 184 == STA/LTA Configuration ==
185
robert 43.1 186 The Short Term Average and Long Term Average Configuration screen.. this is a relic for when people weren't recording continuously. Not advised!
Jack Dent 31.3 187
robert 27.2 188 = LogFile Conversion Script =
189
robert 72.1 190 Both the TSAWR and LPR-200 write logging information as a binary "dat" file which includes GPS time syncs, temperatures, battery power, and position. To convert them into ASCII you can read, use [[THIS PYTHON SCRIPT>>http://auspass.edu.au/field/anusr_log.py]] (current version: 1.42, 02/2026).
robert 27.2 191
robert 28.1 192 = Instrument Response =
robert 27.2 193
robert 60.2 194 Both the TerraSAWR and LPR-200 use the same ADS1281 analog-to-digital converter chip and are designed to have identical instrument response. The ADC (analog to digital) chip in both loggers originally samples at 1024000 Hz and downsamples towards the output data rate via a 5th order SINC filter, then another four FIR filters. If the output is below 250 Hz, a final "pure" /5 decimation is done without any sort of FIR filter (for better or worse!).
robert 1.1 195
robert 62.1 196 In the logger's menu, the user can choose to apply a 2nd stage "sensor gain" by selecting an instrument type in the setup menu. This effectively selects a 10 Vpp (e.g. short period sensors), 20 Vpp, 40 Vpp (most broadband sensors) regime to match the sensor's sensitivity. This has the effect of doubling amplitude from 10v to 20v, or quadrupling from 10v to 40v. If you have set your sensor correctly (and the signal isn't clipped!) you can "correct" this by simply multiplying your data by 0.5 etc. This gain manifests itself in stage 2 in the response information.
197
robert 60.2 198 //(The 600+ Stage 3 SINC coefficients during the initial 1024k > 16k decimation were left off as they slowed down the process x10 and contribute at most 0.3 db amplitude and 0.31 ms phase delay discrepancies, and primarily only to frequencies near the nyquist. If for some reason you want to add this phase manually we can share the parameters with you.)//
robert 28.1 199
robert 62.1 200 You might notice that the response information may come in two versions. The response from our website (see link below) includes the 2nd "sensor gain" stage for clarity. e.g. here is a Trillium Compact 120 & ANU Logger response
201
202 ##Channel Response
203 From M/S (Velocity) to COUNTS ()
204 Overall Sensitivity: 3.95452e+08 defined at 1.000 Hz
205 8 stages:
206 Stage 1: PolesZerosResponseStage from M/S to V, gain: 754.3
207 Stage 2: ResponseStage from V to V, gain: 0.25
208 Stage 3: CoefficientsTypeResponseStage from V to COUNTS, gain: 2.09715e+06
209 Stage 4: FIRResponseStage from COUNTS to COUNTS, gain: 1
210 Stage 5: FIRResponseStage from COUNTS to COUNTS, gain: 1
211 Stage 6: FIRResponseStage from COUNTS to COUNTS, gain: 0.99998
robert 63.1 212 Stage 7: FIRResponseStage from COUNTS to COUNTS, gain: 1##
robert 62.1 213
214
robert 64.1 215 However, if retrieving from AusPass or IRIS, the 2nd "sensor gain" stage is combined with the logger gain. This has no affect, but you may detect that the former Stage 2 V->V ResponseStage has been merged into the Stage 3 gain.
robert 62.1 216
217 ##Channel Response
218 From M/S (Velocity) to COUNTS ()
219 Overall Sensitivity: 3.9546e+08 defined at 1.000 Hz
220 6 stages:
221 Stage 1: PolesZerosResponseStage from M/S to V, gain: 754.3
222 Stage 2: CoefficientsTypeResponseStage from V to COUNTS, gain: 524288
223 Stage 3: FIRResponseStage from COUNTS to COUNTS, gain: 1
224 Stage 4: FIRResponseStage from COUNTS to COUNTS, gain: 1
225 Stage 5: FIRResponseStage from COUNTS to COUNTS, gain: 0.99998
226 Stage 6: FIRResponseStage from COUNTS to COUNTS, gain: 1##
227
228
robert 60.2 229 For the most part, the data logger response essentially flat when the samplerate output is set to 100 Hz or less and for seismological purposes is likely to be impossible to detect below 20 Hz regardless.
robert 28.1 230
robert 71.1 231 Instrument response can be downloaded from IRIS [[Nominal Response Library>>https://ds.iris.edu/ds/nrl/datalogger/anurses]] if need be, or [[directly from us>>http://auspass.edu.au/data/logger_response]] , or by downloading the response of an equivalent sensor at AusPass (e.g. get_stations(level='response') ).
robert 28.1 232
robert 56.1 233
robert 59.1 234 [[Amplitude and phase response for ANU logger at 50 Hz>>image:ANU_50hz_response.png||data-xwiki-image-style-alignment="center" height="356" width="475"]]
robert 56.1 235
robert 59.1 236 [[Amplitude and phase response for ANU logger at 100 Hz>>image:ANU_100hz_response.png||data-xwiki-image-style-alignment="center" height="355" width="473"]]
robert 56.1 237
robert 59.1 238 [[Amplitude and phase response for ANU logger at 250 Hz>>image:ANU_250hz_response.png||data-xwiki-image-style-alignment="center" height="359" width="479"]]
robert 56.1 239
robert 59.1 240 [[Amplitude and phase response for ANU logger at 1000 Hz>>image:ANU_1000hz_response.png||data-xwiki-image-style-alignment="center" height="367" width="489"]]
robert 56.1 241
242
robert 60.2 243 [[Huddle test comparing a Trillium Compact 120 + TerraSAWR vs a Trillium Compact 120 + Nanometrics Centaur (M8.AUANU) at 100 Hz>>image:TC120_ANU_vs_CENTAUR.png||data-xwiki-image-style-alignment="center"]]
robert 28.1 244
robert 55.1 245 = ANU TerraSAWR (Gen 3, FW 3.5a, 2014- current) =
robert 28.1 246
robert 55.1 247 Earliest known model is dated July 2014 (though first deployed in 2019) and our current flagship model. Lightweight and small.
robert 1.1 248
robert 55.1 249 = ANU LPR-200 (Gen 2, FW 2.6a/2.7a, 2011 - current) =
robert 1.1 250
robert 55.1 251 Earliest known model is dated May 2011 (but first deployed November 2012) and still in use today. Potentially capable of housing much larger batteries than the TSAWR due to the larger cavity space.
robert 1.1 252
robert 69.1 253 We have found that the older LPR SD card slots can sometimes fail such that the data cards prematurely "pop out" during recording, which can be catastrophic. This seems to mostly occur in periods of hot weather, but not always. We have developed a small add-on that screws into the nearby serial port that effectively holds the cards in place and will ship included with these loggers. Screw the device in firmly, then **gently** slide over a pre-inserted card until it is flush with the top of the card. It probably won't slide over all the way and that's by design. Take care not to force the depress the card inadvertently (although testing has shown it will still record).
robert 68.1 254
255 [[image:LPR_card_adaptors.jpg||alt="image of the LPR SD-card adapter"]]
256
robert 46.2 257 = ANU "ANUSR" (Gen 1, 2003? - 2012) =
robert 1.1 258
robert 46.2 259 This logger has been retired for a long time and has a different instrument response. It used modular component boards and was powered via an acrylic case of 6 x 6V lantern batteries. There is a somewhat complete one above the CAT lab door if anyone is so inclined to have a look.
robert 1.1 260
261
262 = References =
263
robert 42.1 264 * [[PDF Manual>>http://auspass.edu.au/field/LPR-200_Instruction_Manual.pdf]]
robert 73.1 265
266 [[TSAWR & LPR-200 sensor pinout>>image:ANU_seismometer_connection.png]]
267
268
robert 42.1 269 )))
robert 1.1 270
271 (% class="col-xs-12 col-sm-4" %)
272 (((
273 (% class="box" %)
274 (((
robert 6.1 275 **Contents**
robert 1.1 276
robert 42.1 277 {{toc/}}
robert 6.1 278 )))
279
Jack Dent 17.1 280
robert 45.1 281
robert 46.1 282
robert 46.2 283
robert 48.1 284
robert 49.1 285
robert 50.1 286
robert 50.2 287
Jack Dent 52.1 288
Jack Dent 53.1 289
santi 54.1 290
robert 55.1 291
robert 56.1 292
robert 57.1 293
robert 58.1 294
robert 58.2 295
robert 59.2 296
robert 60.2 297
robert 62.1 298
robert 63.1 299
robert 64.1 300
robert 65.1 301
robert 66.1 302
robert 67.1 303
robert 68.1 304
robert 69.1 305
robert 70.1 306
robert 71.1 307
robert 71.2 308
robert 73.1 309
Jack Dent 73.2 310
Jack Dent 74.1 311
Jack Dent 75.1 312
Jack Dent 76.1 313
Jack Dent 77.1 314
Jack Dent 17.1 315 (% class="box" %)
316 (((
robert 7.1 317 = TerraSAWR Specs =
robert 6.1 318
robert 1.1 319
robert 42.1 320 [[image:Terrasawr 1.jpg]]
Jack Dent 9.1 321
robert 42.1 322 [[image:1704865206800-824.jpg]]
Jack Dent 9.1 323
robert 42.1 324 [[image:1704864886951-793.jpg]]
Jack Dent 9.1 325
robert 11.1 326 |=Size (L x W x H)|30 x 25 x12cm
robert 45.1 327 |=Weight (with battery)| 1.9kg (2.5kg)
robert 14.1 328 |=Battery|(((
robert 46.2 329 Lead-acid 12v, ~~7-9Ah
robert 2.2 330
331 Rechargeable
robert 1.1 332 )))
robert 46.2 333 |=Current Firmware|3.5a (Jan 2025)
Jack Dent 16.1 334
robert 42.1 335 [[image:terrasawr battery.jpg]]
robert 2.2 336 )))
robert 1.1 337
Jack Dent 17.1 338
robert 1.1 339 (% class="box" %)
340 (((
robert 7.1 341 = LPR-200 Specs =
robert 3.1 342
robert 42.1 343 [[image:LPR 1.jpg]]
robert 3.1 344
robert 42.1 345 [[image:LPR 2.jpg]]
Jack Dent 9.1 346
robert 42.1 347 [[image:LPR 3.jpg]]
Jack Dent 9.1 348
robert 11.1 349 |=Size (L x W x H)|42 x 34 x17 cm
robert 45.1 350 |=Weight (no battery)|5.5 kg
robert 14.1 351 |=Battery|(((
robert 46.1 352 NOW: Any 12v battery with tab connections that will fit (ex. the same TSAWR battery)
robert 3.1 353
robert 46.2 354 DISCONTINUED: LiFePo4 140Ah Rechargeable (shown below).
robert 3.1 355 )))
robert 46.2 356 |=Current Firmware|(((
357 2.6a (old GPS modules) / 2.7a
Jack Dent 16.1 358
robert 46.2 359 Jan 2025
360 )))
361
robert 42.1 362 [[image:LPR battery.jpg]]
robert 3.1 363 )))
robert 1.1 364 )))
robert 61.1 365
366
367
robert 62.1 368
robert 63.1 369
robert 64.1 370
robert 65.1 371
robert 66.1 372
robert 67.1 373
robert 68.1 374
robert 69.1 375
robert 70.1 376
robert 71.1 377
robert 71.2 378
robert 73.1 379
Jack Dent 73.2 380
Jack Dent 74.1 381
Jack Dent 75.1 382
Jack Dent 76.1 383
Jack Dent 77.1 384
robert 61.1 385
robert 42.1 386 )))