Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.JackD
1 +XWiki.robert
Content
... ... @@ -12,18 +12,12 @@
12 12  
13 13  = Power Considerations =
14 14  
15 -Both the LPR-200 (or "Low Power Recorder" 200) and TerraSAWR are designed to use as little power as possible.
15 +Both the LPR-200 (or "Low Power Recorder" 200) and TerraSAWR are designed to use as little power as possible, and more or less use the same amount of power.
16 16  
17 17  At 100 Hz and with a GPS cable connected these loggers draw about 220 mW of power once the screen is off (higher sample rates draw more power but only marginally, < 5 mW). Adding a sensor (e.g. a Trillium Compact 120) increases this to approximately 400 mW, or 0.4 volt-amps. So, in theory 7 Ah battery should last about 10 days without a solar panel, but in practice it seems to be a bit closer to 8 which may be due to variability in power drain while in getting GPS locks.
18 18  
19 -For very sunny environments (latitudes < 30) a 20 Volt, 10 Watt solar panel should have no issue keeping these loggers alive over the summer months, and assuming unobstructed skies, should also be fine over winter. **However** when possible a 20 Watt panel works best, especially as they usually cost only $10-20 more these days.
19 +For very sunny environments (latitudes < 30) a 20 Volts 10 Watt solar panel should have no issue keeping these loggers alive over the summer months, and assuming unobstructed skies should also be fine over winter. However there is no harm in using 20 or even a 40 Watt panel, especially for high latitudes, coastal regions, or areas without a full sky view. In theory up to a 60 Watt solar panel is fine, but we don't recommend anything over 40 Watts and that amount of power is already overkill.
20 20  
21 -Measurements of a 20 Watt panel in the Canberra (Aus) area show that in full sun, roughly 0.75 amps are put out. As the LPR-200 and TSAWR loggers step down the solar voltage to 12v, a 20 Watt solar panel realistically provides 9 Watts to a seismic station. Theoretically, 1 hour and 9 minutes of sun per day should keep the battery charge stable. However, in practice, 1 hour and 30 minutes of full sunlight for a 20 Watt solar panel is required to keep these stations at a stable charge (this was tested over a 5 day period under Canberra sun in spring)
22 -
23 -**Note**: It is important to understand 2 main limitations of the solar component of these seismic stations. First, the voltage output of most solar panels will drop significantly if even a small portion of the panel is shaded. This is why 'full' sun is a requirement. Second, 1 hour and 30 minutes of full sun per day will not reliably be achieved each day. Thus, the sunniest possible location for a seismic station is always ideal.
24 -
25 -40 Watt panels can also be used, and may be needed for extreme climates, or areas without a full sky view. 60 Watt panels have also been known to work, but this is now reaching the the limit of what the loggers can reasonably handle for extended periods and are not recommended for long term deploys.
26 -
27 27  (% class="box infomessage" %)
28 28  (((
29 29  Power issues are easy and cheap to solve relative to the cost of your experiment, don't skimp!
... ... @@ -32,12 +32,6 @@
32 32  (% class="wikigeneratedid" %)
33 33  In the case of an LPR, there is a large compartment for housing an internal battery, able to accommodate anything from a 10-30Ah battery. To use a standard lead acid battery with a positive and negative terminal, a 6 pin adaptor must be used. This ensures the voltage from the external power port (pins A and C) connect to the battery and ensure the system actually recharges. (See [[Peripheral Equipment>>doc:Instrumentation.Peripheral Equipment.WebHome]] for a more comprehensive overview of this kind of setup)
34 34  
35 -=== Recharging a logger battery ===
36 -
37 -If a logger stops receiving solar power, the station will drain and the battery may become unusable. Upon servicing a station in the field, the voltage of the 12v lead-acid battery should be assessed while not powering anything. If the voltage is below 10.5v, it is likely not suitable for use in the field and should be replaced with a functional battery. If the voltage is below 12.5v, it is a good idea to replace it with a fully charged 12v battery (though if this is not possible, it is ok to leave this battery so long as the station was alive on arrival.
38 -
39 -If a 12v lead-acid battery is collected from the field and is still above 10.5v, recharge it fully using a 12v battery charger. These are easily acquired at auto or electronics store. Recharging a 12v lead-acid battery with a capacity of about 10Ah should take a couple of hours (~~2.5hrs).
40 -
41 41  = GPS Considerations =
42 42  
43 43  GPS is required for the data to have accurate timestamps. A standard 3-5V 1575.42 Mhz coaxial "dongle" antenna works fine and can be found for relatively cheap (e.g. [[https:~~/~~/www.elecbee.com/en-3555-gps-antenna-bnc-male-for-garmin-gps-120120xl125-sounder-with-cable-2m) >>https://www.elecbee.com/en-3555-gps-antenna-bnc-male-for-garmin-gps-120120xl125-sounder-with-cable-2m]]. Anything that locates satellites and gives you a GPS-synced time works!
... ... @@ -70,7 +70,7 @@
70 70  
71 71  (% class="box warningmessage" %)
72 72  (((
73 -NOTE: the 2 at the very end is for "RECORD ON RESTART". The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume. (# of blank spaces before this doesn't matter)
61 +NOTE: the 2 at the very end is for "RECORD ON RESTART". The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume. (# of blank spaces before this doesn't matter)
74 74  )))
75 75  
76 76  (% class="wikigeneratedid" id="HTheformatforTSAWRloggersisshorter:" %)
... ... @@ -187,7 +187,7 @@
187 187  
188 188  = LogFile Conversion Script =
189 189  
190 -Both the TSAWR and LPR-200 write logging information as a binary "dat" file which includes GPS time syncs, temperatures, battery power, and position. To convert them into ASCII you can read, use [[THIS PYTHON SCRIPT>>http://auspass.edu.au/field/anusr_log.py]] (current version: 1.42, 02/2026).
178 +Both the TSAWR and LPR-200 write logging information as a binary "dat" file which includes GPS time syncs, temperatures, battery power, and position. To convert them into ASCII you can read, use [[THIS PYTHON SCRIPT>>http://auspass.edu.au/field/anusr_log.py]].
191 191  
192 192  = Instrument Response =
193 193  
... ... @@ -228,7 +228,7 @@
228 228  
229 229  For the most part, the data logger response essentially flat when the samplerate output is set to 100 Hz or less and for seismological purposes is likely to be impossible to detect below 20 Hz regardless.
230 230  
231 -Instrument response can be downloaded from IRIS [[Nominal Response Library>>https://ds.iris.edu/ds/nrl/datalogger/anurses]] if need be, or [[directly from us>>http://auspass.edu.au/data/logger_response]] , or by downloading the response of an equivalent sensor at AusPass (e.g. get_stations(level='response') ).
219 +Instrument response can be downloaded from IRIS [[Nominal Response Library>>https://ds.iris.edu/ds/nrl/]] if need be, or [[directly from us>>http://auspass.edu.au/data/logger_response]] , or by downloading the response of an equivalent sensor at AusPass (e.g. get_stations(level='response') ).
232 232  
233 233  
234 234  [[Amplitude and phase response for ANU logger at 50 Hz>>image:ANU_50hz_response.png||data-xwiki-image-style-alignment="center" height="356" width="475"]]
... ... @@ -250,7 +250,7 @@
250 250  
251 251  Earliest known model is dated May 2011 (but first deployed November 2012) and still in use today. Potentially capable of housing much larger batteries than the TSAWR due to the larger cavity space.
252 252  
253 -We have found that the older LPR SD card slots can sometimes fail such that the data cards prematurely "pop out" during recording, which can be catastrophic. This seems to mostly occur in periods of hot weather, but not always. We have developed a small add-on that screws into the nearby serial port that effectively holds the cards in place and will ship included with these loggers. Screw the device in firmly, then **gently** slide over a pre-inserted card until it is flush with the top of the card. It probably won't slide over all the way and that's by design. Take care not to force the depress the card inadvertently (although testing has shown it will still record).
241 +We have found that the older LPR SD card slots can sometimes fail such that the data cards prematurely "pop out" during recording, which can be catastrophic. This seems to mostly occur in periods of hot weather, but not always. We have developed a small add-on that screws into the nearby serial port that effectively holds the cards in place and will ship included with these loggers. Screw the device in firmly, then **gently** slide over a pre-inserted card. Take care not to force the depress the card inadvertently (although testing has shown it will still record).
254 254  
255 255   [[image:LPR_card_adaptors.jpg||alt="image of the LPR SD-card adapter"]]
256 256  
... ... @@ -262,10 +262,6 @@
262 262  = References =
263 263  
264 264  * [[PDF Manual>>http://auspass.edu.au/field/LPR-200_Instruction_Manual.pdf]]
265 -
266 -[[TSAWR & LPR-200 sensor pinout>>image:ANU_seismometer_connection.png]]
267 -
268 -
269 269  )))
270 270  
271 271  (% class="col-xs-12 col-sm-4" %)
... ... @@ -302,16 +302,6 @@
302 302  
303 303  
304 304  
305 -
306 -
307 -
308 -
309 -
310 -
311 -
312 -
313 -
314 -
315 315  (% class="box" %)
316 316  (((
317 317  = TerraSAWR Specs =
... ... @@ -372,15 +372,5 @@
372 372  
373 373  
374 374  
375 -
376 -
377 -
378 -
379 -
380 -
381 -
382 -
383 -
384 -
385 385  
386 386  )))
ANU_seismometer_connection.png