Changes for page ANU Seismic Data Loggers

Last modified by robert on 2025/06/27 16:42

From version 25.1
edited by robert
on 2024/01/23 12:39
Change comment: There is no comment for this version
To version 24.1
edited by robert
on 2024/01/23 12:37
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -18,7 +18,7 @@
18 18  
19 19  The format for "ANUSRSetup.txt" / LPR200s will be a single line of text that looks like this:
20 20  
21 -{{{XXX195G0100010034864 2 }}}
21 +XXX195G0100010034864      2
22 22  
23 23  corresponding to site XXX19,
24 24   network 5G,
... ... @@ -26,16 +26,17 @@
26 26   "gps interval" 01 (once per hour~-~- don't change),
27 27   start mode (always 0),
28 28   stop mode (always 0),
29 - seismometer type (3 = broadband (+/- 20V), 1 = shortperiod (+/- 5V),
29 + seismometer type (3 = broadband, 1 = shortperiod),
30 30   and seismometer serial number (4864).
31 31   the 2 at the very end is for "start recording on power"
32 - and is vital to have this set so the logger will resume recording on its own if it experiences a power cycle. (# of blank spaces before this doesn't matter)
32 + and is vital to have this set so the logger will
33 + resume recording on its own if it experiences a power cycle.
33 33  
34 34  the TSAWRs are a bit shorter as they don't have an option to set the "gps sync"
35 35  
36 -{{{SITE1XX02500039999 2}}}
37 +SITE1XX02500039999      2
37 37  
38 -which would set up a Trillium Compact with serial 9999 named XX.SITE1 at 250hz sample rate.
39 +which would set up a trillium with serial 9999 named XX.SITE1 at 250hz sample rate.
39 39  
40 40  
41 41  Here is a general guideline for how much data you can expect to fit on a card: